Medical Action Industries, Inc. 306 recalls Kit: Epidural/Pain TQC 20/Cs
Kits containing Sodium Chloride ampules subject to another recall are recalled due to potentially compromised sterility.
- Nerve Block Tray
- Nonconforming Material/Component
The FDA has posted 9 enforcement recalls of nerve block tray devices since 2025, including 3 Class I (most serious) recalls. Most recent report: Sep 16, 2026.
Kits containing Sodium Chloride ampules subject to another recall are recalled due to potentially compromised sterility.
Windstone Medical Packaging recalled specific items and lots of AMS convenience kits containing Elevaris Lidocaine HCl injection USP. Elevaris Medical Devices Inc.…
The kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd.…
Medline kits were distributed containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USP, which was subsequently recalled due to quality issues, including microbiology testing data integrity concerns and repor…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles;…
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the s…
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles;…
Source: U.S. FDA openFDA device enforcement reports, updated continuously. Device type per the FDA's RES classification (“Nerve Block Tray”). Informational only — verify against the FDA before acting.