Skip to content
RecallWatchMedical Device Safety
Manufacturer

Becton, Dickinson and Company, BD Bio Sciences recalls

The FDA has posted 9 enforcement recalls from Becton, Dickinson and Company, BD Bio Sciences since 2025, none of them Class I (most serious). Most recent report: Aug 13, 2025.

Class II: 9
Class IIOngoingZ-2079-2025

Becton, Dickinson And Company, BD Bio Sciences recalls REF: 666289

Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.

  • Differential Cell Counter
  • Process control
Becton, Dickinson and…CAJul 9, 2025
Class IIOngoingZ-2078-2025

Becton, Dickinson And Company, BD Bio Sciences recalls REF: 335775

Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.

  • Differential Cell Counter
  • Process control
Becton, Dickinson and…CAJul 9, 2025
Class IIOngoingZ-0745-2025

Becton, Dickinson And Company, BD Bio Sciences recalls FACSLyric 2L4C Instrument US

On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric f…

  • Flow Cytometric Reagents And Accessories.
  • Nonconforming Material/Component
Becton, Dickinson and…CAJan 1, 2025
Class IIOngoingZ-0749-2025

Becton, Dickinson And Company, BD Bio Sciences recalls FACSLyric 3L12C Instrument RUO

On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric f…

  • Flow Cytometric Reagents And Accessories.
  • Nonconforming Material/Component
Becton, Dickinson and…CAJan 1, 2025
Class IIOngoingZ-0748-2025

Becton, Dickinson And Company, BD Bio Sciences recalls FACSLyric 3L10C Instrument US-IVD

On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric f…

  • Flow Cytometric Reagents And Accessories.
  • Nonconforming Material/Component
Becton, Dickinson and…CAJan 1, 2025
Class IIOngoingZ-0747-2025

Becton, Dickinson And Company, BD Bio Sciences recalls FACSLyric 3L8C Instrument US-IVD

On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric f…

  • Flow Cytometric Reagents And Accessories.
  • Nonconforming Material/Component
Becton, Dickinson and…CAJan 1, 2025

Firm name as filed with the FDA: “Becton, Dickinson and Company, BD Bio Sciences”. Related firm entities may file under separate names. Source: U.S. FDA openFDA device enforcement reports. Informational only — verify with the FDA before acting.