Skip to content
RecallWatchMedical Device Safety
Manufacturer

I.T.S. GmbH recalls

The FDA has posted 17 enforcement recalls from I.T.S. GmbH since 2025, none of them Class I (most serious). Most recent report: Apr 1, 2026.

Class II: 17
Class IIOngoingZ-1593-2026

I.T.S. GmbH recalls Claviculaplate with Angular Stability

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

  • Bone Fixation Plate
  • No Marketing Application
I.T.S. GmbHApr 1, 2026
Class IIOngoingZ-1606-2026

I.T.S. GmbH recalls HCS - Headless Compression Screw System

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

  • Bone Fixation Screw
  • No Marketing Application
I.T.S. GmbHApr 1, 2026

Firm name as filed with the FDA: “I.T.S. GmbH”. Related firm entities may file under separate names. Source: U.S. FDA openFDA device enforcement reports. Informational only — verify with the FDA before acting.