Ortho-Clinical Diagnostics, Inc. recalls

The FDA has posted 13 enforcement recalls from Ortho-Clinical Diagnostics, Inc. since 2025, none of them Class I (most serious). Most recent report: Sep 24, 2025.

Class II: 13
Class IIOngoingZ-2335-2025

Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Ca Slides

Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay cal…

  • Calcium Azo Dye
  • Under Investigation by firm
Ortho-Clinical Diagnos…NYAug 27, 2025
Class IIOngoingZ-2122-2025

Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products CHE Slides

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution;…

  • Cholinesterase Colorimetry
  • Under Investigation by firm
Ortho-Clinical Diagnos…NYJul 23, 2025
Class IIOngoingZ-2039-2025

Ortho-Clinical Diagnostics, Inc. recalls Brand Name: VITROS

a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration da…

  • For Clinical Use Chemistry (Photometric, Discrete) Analyzer
  • Software design
Ortho-Clinical Diagnos…NYJul 9, 2025