Tandem Diabetes Care, Inc. recalls

The FDA has posted 7 enforcement recalls from Tandem Diabetes Care, Inc. since 2025, including 4 Class I (most serious) recalls. Most recent report: Apr 1, 2026.

Class I: 4Class II: 3
Class IIOngoingZ-0232-2026

Tandem Diabetes Care, Inc. recalls t:slim X2 insulin pump with Interoperable Technology

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpec…

  • Alternate Controller Enabled Insulin Infusion Pump
  • Software change control
Tandem Diabetes Care,…CAOct 29, 2025
Class IOngoingZ-2343-2025

Tandem Diabetes Care, Inc. recalls Battery Operated infusion pumps. t:slim X2 G5

Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a…

  • Alternate Controller Enabled Insulin Infusion Pump
  • Nonconforming Material/Component
Tandem Diabetes Care,…CASep 3, 2025
Class IOngoingZ-1519-2025

Tandem Diabetes Care, Inc. recalls t:slim X2 Insulin Pump with Interoperable Technology

A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Gl…

  • Alternate Controller Enabled Insulin Infusion Pump
  • Software Design Change
Tandem Diabetes Care,…CAApr 23, 2025
Class IOngoingZ-1520-2025

Tandem Diabetes Care, Inc. recalls Tandem Mobi Insulin Pump with Interoperable Technology

A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Gl…

  • Alternate Controller Enabled Insulin Infusion Pump
  • Software Design Change
Tandem Diabetes Care,…CAApr 23, 2025

Firm name as filed with the FDA: “Tandem Diabetes Care, Inc.”. Related firm entities may file under separate names. Source: U.S. FDA openFDA device enforcement reports. Informational only — verify with the FDA before acting.