CareFusion 303, Inc. recalls BD Alaris Pump Infusion Sets for the following reference numbers: C24101E
Reason for recall
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- BD Alaris Pump Infusion Sets for the following reference numbers: C24101E, Amber Tubing, 15 Micron Filter, 0.2 Micron Filter, SmartSite Y-Site 10015414, 180 Micron Filter 2447-0007, Ball Valve SmartSite Port (Burette) 2 SmartSites Y-Sites 24201-0007, 2 SmartSite Y-sites 10012645, Back Check Valve, Non-vented, SmartSite Y-site 2426-0500, Back Check Valve, 3 SmartSite Y-sites 2450-0500, Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSite Y-sites 11522558, Ball Valve, Back Check Valve, 3 SmartSite Y-sites 2410-0500, 2 SmartSite Y-Sites 2232-0007, 0.2 Micron Filter 11613191, 0.2 Micron Filter, Ball Valve, SmartSite" Port (Burette), 2 SmartSite Y-sites 10012144, 3 SmartSite Y-Sites, Check Valve 10013034, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 4-way Stopcock 10013037, Non-Vented, 180 Micron Filter, Low Sorbing Tubing Segment, 1 SmartSite, Y-Site 11582773, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 3-way StopcockUDI C24101E28 affected lotsC24101E1088540323552810015414108854032339992447-00071088540323507824201-000710885403476341
+20 more
10012645108854032337532426-0500761320302099211522558108854032323052410-050076132030124922232-0007108854034639691161319110885403276026100121447613203021159100130341088540323231210013037108854032337841158277310885403239588
What the firm is doing
On December 12, 2025, CareFusion 303, Inc. (BD) issued a updated "Urgent Medical device Recall Notice via Email and physical mail. The updated notice includes important information on infusion sets as well as labeling updates. Previous notifications were sent on July 8,2025, July 17, 2025, September 11, 2025 and September 12, 2025. CareFusion ask consignees to take the following actions: 1. Review your inventory for the catalog numbers included in the updated letter. Remove and discard any remaining inventory per your facility guidelines. 2.To receive credit or replacement product, please return the completed customer response form attached to this notice. 3.Please contact your BD Clinical Consultants and Account Executives for alternative sets information. 4. Review the updated sets performance data included in the letter. 5. Circulate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 6. Distributors should provide a copy of this notice to all customers who may have purchased an Alaris Pump Module. If you require further assistance, please contact: BD Technical Support at Phone: 888-812-3229; Phone hours: 6:00am to 5:00pm PT Monday Friday and Email: DL-US-INF-TechSupport@bd.com.
DistributionShow detailsHide
US: Nationwide OUS: United Arab Emirates Australia Belgium Bahrain Botswana Canada Switzerland China France United Kingdom Gibraltar Israel India Kuwait Mexico Malaysia Netherlands New Zealand Philippines Pakistan Qatar Saudi Arabia Singapore Taiwan, Province of China South Africa
A reasonable probability that use of the product will cause serious adverse health consequences or death.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-0035-2026
- FDA 510(k) clearance · K022209The device's official FDA premarket clearance record
- FDA device classification · FPAOfficial FDA classification for this device type
- CFR regulation · 21 CFR 880.5440The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find CareFusion 303, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class I recalls
Spectra Medical Devices, LLC recalls Brand Name: Elevaris/Spectra Medical Devices Product Name: Sodium Chloride Injection
Manufacturing firm, Huons, was cited for issues regarding product quality and sterility assurance. If the product sterility is compromised, patients may be exposed to microorganisms, endotoxins, or foreign matter.…
- Vascular Access Flush Saline
- Nonconforming Material/Component
Baxter Healthcare Corporation recalls Baxter Duo-Vent Solution Set and Male Luer Lock Adapter
Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position.
- Intravascular Administration Set
- Under Investigation by firm
Boston Scientific Neuromodulation Corporation recalls INFINION CX LEAD KIT
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, i…
- Totally Implanted For Pain Relief Spinal-Cord Stimulator
- Device Design