Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-0067-2027

Custom Interventional Pain Management, LLC recalls Brand Name: Ferndale Laboratories

Custom Interventional Pain Management, LLCSan Antonio, TX, United StatesReported Oct 7, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Due to manufacturing issues with the liquid skin adhesive containing in surgical convenience trays/kits.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Brand Name: Ferndale Laboratories; Medline; Custom Interventional Pain Management CIPM Product Name: Mastisol Skin Adhesive Model/Catalog Number: 24161B (Ferndale and Medline); Tray 17025 17g continuous PBN tray (CIPM) Software Version: n/a Product Description: Skin Adhesive in a narrow plastic sterilized tube. Mastisol is a skin glue often used during surgical procedures. Continuous epidural trays are designed for use by anesthesiologists and radiologists in the administration of regional anesthesia. Component: surgical kits

Lot / code information

Lot #
Code: CIPM product / reorder #17025; Description: 17g continuous PBN tray
Part / Item #
110026; Received
Lot #
27022; Affected
Lot #
1225271; Expiration Date: 01/31/2028; DHR or Batch Record #64591
UDI
*+B87517LIDO/$$20131281225271T* UDI- DI #: B87517LID0 with Lidocaine

What the firm is doing

On 01/15/2026, a customer notification letter was email to a single customer informing them that, the MASTISOL Liquid adhesive shipped in kits in lot 1225271, P.O.#2513972 was exposed to a double sterilization cycle by Ferndale Laboratories, Inc. which may contribute to the butyrate tubing being brittle and cracking during actuation. The customer was instructed to: 1. lmmediately examine your inventory and quarantine Lot 1225271. 2. Return all units of MASTISOL in Lot 1225271 to: CIPM 12460 Network Blvd. Suite 106 San Antonio, TX 78249 3. Complete the enclosed Return Materials Authorization Form QF-85-05-02 and return to denise@cipmtrays.com or quality@cipmtrays.com. For questions - call Toll Free 1-888-392-4514

DistributionShow details

U.S. Nationwide distribution in the state of TX.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-3334-2026

Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…

  • Bone Fixation Plate
  • Process control
Wright Medical Technol…TNOct 7, 2026