Class IIOngoingZ-0124-2026

B BRAUN MEDICAL INC recalls Design Options

B BRAUN MEDICAL INCAllentown, PA, United StatesReported Oct 15, 2025 · 9 months ago
Legal News Analyst ·

Reason for recall

Potential for the lid of the catheter connector to be in the incorrect position.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Design Options, Pain Management Tray. Product Code: SESK. Product Descriptions: 1. SESK PARKLAND MEMORIAL FX CSE TRAY. 2. SESK OWENSBORO DAVIESS HOSP 20058814. 3. SESK MEMORIAL HEALTH SYSTEM SPINAL/EPID. 4. SESK ST. JOHN'S MERCY MED CTR SPINAL/EPID. 5. SESK MAIMONIDES HOSPITAL SPINAL/EPIDURAL. 6. SESK OU MEDICAL CENTER - SPINAL/EPIDURAL. 7. SESK ELLIS HOSPITAL - SPINAL EPIDURAL. 8. STONY BROOK HSP SPINAL EPIDRAL TRAY SESK. 9. SESK UAB SPINAL EPIDURAL. 10. SESK CREIGHTON UNIV BMH SOINAL EPIDURAL. 11. SESK, BSW SPINAL EPIDURAL TRAY. 12. LYNDON B JOHNSON HOSPITAL SPINAL EPID.

Lot / code information

REF
530029; Product Code: SESK
UDI
(Primary): 04046964305245
UDI
(Unit of Use): 04046964305238
Lot #
0062012031. 2
REF
530048; Product Code: SESK
UDI
(Primary): 04046964305313
UDI
(Unit of Use): 04046964305306
Lot #
0062006298. 3
REF
530155; Product Code: SESK
UDI
(Primary): 04046964306211
Show 38 more code fields
UDI
(Unit of Use): 04046964306204
Lot #
0062012054. 4
REF
530159; Product Code: SESK
UDI
(Primary): 04046964669286
UDI
(Unit of Use): 04046964669279
Lot #
0062012025. 5
REF
530169; Product Code: SESK
UDI
(Primary): 04046964808920
UDI
(Unit of Use): 04046964808913
Lot #
0062012035. 6
REF
530171; Product Code: SESK
UDI
(Primary): 04046964950391
UDI
(Unit of Use): 04046964950384
Lot #
0062006303. 7
REF
530180; Product Code: SESK
UDI
(Primary): 04046955156443
UDI
(Unit of Use): 04046955156436
Lot #
0062000738. 8
REF
530197; Product Code: SESK
UDI
(Primary): 04046955619337
UDI
(Unit of Use): 04046955619320
Lot #
0062003710. 9
REF
530194; Product Code: SESK
UDI
(Primary): 4046955589388
UDI
(Unit of Use): 4046955589371
Lot #
0062014360. 10
REF
530199; Product Code: SESK
UDI
(Primary): 04046955619375
UDI
(Unit of Use): 04046955619368
Lot #
0062001876. 11
REF
530160; Product Code: SESK
UDI
(Primary): 04046964813993
UDI
(Unit of Use): 04046964813986
Lot #
0062011011. 12
REF
530184; Product Code: SESK
UDI
(Primary): 04046955156528
UDI
(Unit of Use): 04046955156511
Lot #
0062003228. Expiration Date: 07/31/2026

What the firm is doing

B Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distributed, notify customers, post the notification where affected units are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form." Additionally, if consignees have identified affected units in inventory, they may order one case (50 pieces) of affected sterilized single use catheter connectors to have on hand.

DistributionShow details

Nationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.