Class IIOngoingZ-0274-2026

SEDECAL SA recalls Model Number SM-40HF-B-D-C

SEDECAL SAAlgete, SpainReported Oct 29, 2025 · 8 months ago
Legal News Analyst ·

Reason for recall

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Model Number SM-40HF-B-D-C; 40KW 80C, Mobile X-ray system

Lot / code information

Serial #
G39822, G62662, G62564, G61702, G60067, G38060

What the firm is doing

Sedecal emailed their distributors on 06/10/2025 to remind them the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning. on 09/19/2025 the firm issued an Important Safety Notice to its direct consignees via email. This notice initiates Phase 2 of the recall strategy expanding the scope and providing information on stickers to cover gaps around the fine positioning control buttons and a DMC software update that stops unit movement if triggered by a short cicuit or prolonged button press with instructions. Consignees will be contacted by a Sedecal Customer support Service to coordinate the shipment of the necessary conversion kits (quantity and delivery address) to allow upgrading the affected units.

DistributionShow details

US: CA, IL, and NJ

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.