Class IIOngoingZ-0623-2026

B Braun Medical Inc recalls CARESAFE utilized in gravity IV administration sets and pump administration sets utiliz…

B Braun Medical IncBethlehem, PA, United StatesReported Dec 3, 2025 · 7 months ago
Legal News Analyst ·

Reason for recall

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

CARESAFE utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. CARESAFE IV ADMIN SET w/AIRSTOP, 104 IN.; Catalog Number: 354302. 2. CARESAFE ADMIN SET w/15um FILTER, 104 IN; Catalog Number: 354303. 3. CARESAFE IV ADMIN SET w/AIRSTOP, 114 IN.; Catalog Number: 354304. 4. CARESAFE ADMIN SET w/15um FILTER, 114 IN; Catalog Number: 354305. 5. CARESAFE IV ADMIN SET w/AIRSTOP, 110 IN.; Catalog Number: 354306. 6. CARESAFE ADMIN SET w/15um FILTER, 110 IN; Catalog Number: 354307. 7. CARESAFE" IV Administration Set with AirStop Filter, Not made with PVC or DEHP, 2 CARESITE¿ Injection Sites, 2 ULTRAPORT¿ High-FlowFour-Way Stopcocks; CARESAFE ANESTH SET W/AIRSTOP, 134 IN.; Catalog Number: 354308. 8. CARESAFE ANESTH SET w/15um FILTER,134IN.; Catalog Number: 354310.

Lot / code information

Catalog #
354302; Primary
UDI
04046955259946; Unit of Dose
UDI
04046955259939. 2
Catalog #
354303; Primary
UDI
04046955052424; Unit of Dose
UDI
04046955052417. 3
Catalog #
354304; Primary
UDI
04046955259960; Unit of Dose
UDI
04046955259953. 4
Catalog #
354305; Primary
Show 14 more code fields
UDI
04046955052448; Unit of Dose
UDI
04046955052431. 5
Catalog #
354306; Primary
UDI
04046955259984; Unit of Dose
UDI
04046955259977. 6
Catalog #
354307; Primary
UDI
04046955052462; Unit of Dose
UDI
04046955052455. 7
Catalog #
354308; Primary
UDI
04046955260003; Unit of Dose
UDI
04046955259991. 8
Catalog #
354310; Primary
UDI
04046955052929; Unit of Dose
UDI
04046955052905

What the firm is doing

B Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administration process without the need for the modified workflow. Consider priming all IV administration sets and pump administration sets with normal saline or other standard solutions prior to priming with expensive medications to avoid loss of expensive medications or occupational exposure to hazardous medications in the event the set is unable to be primed. If difficulty priming is encountered, obtain a replacement device and file a customer complaint with B. Braun. Additionally, consignees were requested to ensure all pertinent employees and customers are informed, post the notification where affected products are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form."

DistributionShow details

Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.