Class IOngoingZ-0711-2025

Maquet Cardiovascular, LLC recalls VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indica…

Maquet Cardiovascular, LLCWayne, NJ, United StatesReported Jan 15, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indicated for use in minimally invasive surgery allowing access for vessel harvesting
    UDI-DI 00607567700406.Model Number

What the firm is doing

Getinge notified consignees on 12/09/2024 via Urgent Medical Device Correction letter posted digitally on Getinge.com and sent to consignees via letter. Consignees were not instructed to return affected units, but instead to review the IFU information provided with the letter to minimize over-delivery of energy and mitigate risks. Getinge instructed consignees to notify customers if further distributed, as well as requested to complete and return the MEDICAL DEVICE CORRECTION RESPONSE FORM. Training to reinforce precautions and appropriate use of the device is expected to be deployed and communicated to all customers by end of January 2025. UPDATE: On 01/31/2025 an updated URGENT MEDICAL DEVICE CORRECTION letter was sent to customers. Consignees were instructed to review the IFU information provided in Appendix A of the letter and consider the listed areas to mitigate the risks. Getinge has created a training video, accessible via a QR code or link, to guide users on the proper use of the device and reinforce essential precautions. Consignees were instructed to notify customers if product was further distributed. Additionally, consignees were instructed to complete, sign and return the provided MEDICAL DEVICE CORRECTION RESPONSE FORM. UPDATE: On 07/11/2025 an updated Urgent Medical Device Correction Update letter was posted on Getinge.com and sent to consignees via letter. The updated communication informs users that the Instructions for Use (IFU) for the Vasoview Hemopro 2 EVH System has been revised to address the bent or detached heater wire and silicone peeling or detaching from the Jaws of the Harvesting Tool. Consignees were instructed to review the updated IFU, view the training video linked in the letter, complete and return the MEDICAL DEVICE CORRECTION RESPONSE FORM, and notify customers if product was further distributed.

DistributionShow details

Nationwide distribution. International distribution to Australia, Austria, Bahrain, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Guatemala, Hong Kong, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Netherlands, Netherlands, New Zealand, Oman, Panama, Puerto Rico, Philippines, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2624-2026

Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…

  • Pulmonary Valve Prosthesis Percutaneously Delivered
  • Process control
Medtronic Heart Valves…CAJul 8, 2026
Class IOngoingZ-2622-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026
Class IOngoingZ-2621-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026