PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234
Reason for recall
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number:722234Model Number
What the firm is doing
On October 2, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions that should be taken by the customer/user in order to prevent risks for patients a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the instructions below. b. If you experience inconsistent (not smooth) or no motorized longitudinal movement of the FlexArm stand and you receive (or have received) the error messages described in Section 1 of this Medical Device Correction letter, please report the issue to Philips. c. If the FlexArm s longitudinal motorized movement stops during a procedure, position the C-arm manually by using the handgrips and brake controls on both sides of the FlexArm stand or reposition the table in longitudinal position if the available distance is sufficient. d. As part of the preventative maintenance cycle Philips will clean the rails, friction wheel and bearings as indicated in the Preventive Maintenance Update attached in Appendix A. o Keep a copy of this Preventative Maintenance Manual Update with your current manual. o If you do not use Philips to perform the preventative maintenance on your system, provide a copy of the Preventive Maintenance Manual Update to your qualified and authorized service provider. e. Circulate this Urgent Medical Device Correction Letter to all users so that they are aware of the issue and follow the instructions above. Keep this Urgent Medical Device Correction with the documentation of the system until Philips corrects your system. f. Complete and return the attached response (page 05) form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction and understanding of the issue and actions required. Actions planned by Philips Image Guided Therapy Systems to correct the problem Philips will replace the existing longitudinal bearings on the affected Azurion 7M20 systems with FlexArm stand with re
DistributionShow detailsHide
Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY, OH, OK, PA, RI, TX, UT, VA, WA, WI, WV, WY; International: Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Netherlands, Norway, Philippines, Poland, Romania, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Vietnam;
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-0719-2026
- FDA 510(k) clearance · K200917The device's official FDA premarket clearance record
- FDA device classification · OWBOfficial FDA classification for this device type
- CFR regulation · 21 CFR 892.1650The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) BASIC PACK
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Eye Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) KIT
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Orthopedic Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) CHI STD NEURO CRANI OPT 2
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Neurological Tray
- Nonconforming Material/Component