Abbott Vascular Inc recalls 20/30 Priority Pack Accessory Kit/.096 RHV

Abbott Vascular IncTemecula, CA, United StatesReported Jan 1, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • 20/30 Priority Pack Accessory Kit/.096 RHV, REF: 1000186, or use during vascular procedures in conjunction with interventional and / or diagnostic devices (e.g., balloon dilatation catheters, atherectomy devices, stent delivery systems, intravascular ultrasound devices)UDI-DI
    UDI-DI 08717648013614
    4 affected lots
    60538750605387516053911360545329

What the firm is doing

On 11/6/2024, recall notices were distributed to customers asking them to take the following actions: 1) Stop using affected devices. 2) Complete and return the acknowledgement form. 3) Return unused devices to the firm. 4) Share this notification with relevant personnel in your organization. 5) If you have further distributed/transferred the affected products, notify those customers. If you have any questions contact firm customer service at 800-227-9902

DistributionShow details

US: IA, MD, DC, IN, CA, NV, TX, CT, GA, MS, NY, OH, FL, NC, MI, KY, PA, VA, IL, OK, TN, UT, WA, WI, MT, OR, LA, HI, NE, AR, MO, MA, KS, ID, AL, AZ, PR, CO, SC. OUS: IN, KR, SG, TH, VN, AR, BR, CA, CO, PA, UY, BE, CZ, DE, ES, FR, GB, IT, PT, RE, SE, SK, TM, JP.