Baxter Healthcare Corporation recalls Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6)
Reason for recall
Devices were identified as released after repair without full testing being performed, which includes flow testing.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6), Product Code 35700BAX
Lot / code information
- GTIN
- 00085412091570
- Serial #
- 779515
What the firm is doing
Customers were contacted via telephone beginning 11/19/24. Baxter Healthcare is calling to notify you of a potential issue with the Spectrum infusion pump. One of your devices has been identified as released after repair without full testing being performed, which includes flow testing. Baxter is requesting the return of the impacted infusion pump to perform all tests required and confirm that the infusion pump performs as expected. As part of this process, we need to make you aware of the potential hazard associated with flow related issues. Flow issues may cause excessive therapy, free-flow condition, insufficient therapy or interruption of therapy. These may lead to serious adverse health consequences, including permanent injury and death. Send signed form to Global CORP Service Quality Field Actions Global_CORP_Service_Quality_Field_Actions@baxter.com
DistributionShow detailsHide
US Nationwide distribution in the states of AL, IN.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-0775-2025
- FDA 510(k) clearance · K133801The device's official FDA premarket clearance record
- FDA device classification · FRNOfficial FDA classification for this device type
- CFR regulation · 21 CFR 880.5725The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Baxter Healthcare CorporationSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm