Skip to content
RecallWatchMedical Device Safety
Class IICompletedZ-0839-2025

Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMING

Augustine Temperature Management, LLCMinnetonka, MN, United StatesReported Jan 22, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) Model # WC71, 2) Model # WC71R (refurbished); used in conjunction with Warming Blankets and Mattresses.

Lot / code information

UDI
00850011479018
Serial #
23060144, 22030085, 22030087, 22030088, 23060138, 23060139, 22030075, 22030076, 22030077, 22030079, 22030080, 22030081, 22030082, 22030083, 22030084, 23060141; 2)
UDI
00850011479285
Serial #
20080002

What the firm is doing

Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023. The firm did not perform any notification to its consignees that corrections would be implemented.

DistributionShow details

Worldwide

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.