Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-0899-2026

Medline Industries, LP recalls Curad Quick Strip Fabric Bandages Containing 30 Bandages of 3/4" x 3"

Medline Industries, LPNorthfield, IL, United StatesReported Dec 17, 2025 · 7 months ago
Legal News Analyst ·

Reason for recall

Medline has identified that all lots of the Curad Touch-Free and Quick-Strip Bandages may exhibit open seals in the packaging which may compromise the sterility of the product.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Curad Quick Strip Fabric Bandages Containing 30 Bandages of 3/4" x 3", Medline Item Number CUR1930P

Lot / code information

UDI/DI 10888277702776 (each), 00888277702779 (box), 20888277702773 (case); ALL LOTS

What the firm is doing

Medline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/27/2025 via email and USPS first class mail. The notices explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-220 Recall Code: 3.Your account will receive credit once the response form is submitted. 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

DistributionShow details

US, Guam, Canada, United Arab Emirates

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls