Class IOngoingZ-0914-2026

GE Medical Systems China Co., Ltd. recalls GE Healthcare Carestation anesthesia system

GE Medical Systems China Co., Ltd.Wuxi, ChinaReported Dec 24, 2025 · 6 months ago
Legal News Analyst ·

Reason for recall

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 750 A1, REF 1012-9750-000; Carestation 750 A2, REF 1012-9750-002; Carestation 750c A1, REF 1012-9755-000; Products for which affected PMB can be used as a spare part: Carestation 750c A2, REF 1012-9755-002

Lot / code information

REF
1012-9750-000:
GTIN
00840682145596
REF
1012-9750-002:
GTIN
00840682146470
REF
1012-9755-000:
GTIN
00840682146425
REF
1012-9755-002:
GTIN
00840682146463

What the firm is doing

GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 11/14/2025 via letter using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken by Customer/User: "Pending correction by GE HealthCare, you can continue to use the Anesthesia machine with affected power management boards by following the instructions below: 1. Always ensure the device has a secure connection to an AC mains power source. 2. If there is a loss of AC mains power to the system leading to an unexpected system shutdown: - Promptly initiate ventilation using a self-inflating bag connected to an oxygen source. - Assess oxygenation via pulse oximetry. - Because volatile anesthetic agent delivery may transiently be disrupted, supplement with or transition to intravenous anesthetics as needed. - Following system reboot, the system will enter pre-use check. Press "Start Anesthesia or Start Case and then select the "Bypass" button to bypass the checkout. Proceed to selecting the ventilation parameters and volatile agent concentration appropriate for the patient. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to RECALL.FMI34143@gehealthcare.com or submit the acknowledgment using survey." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

DistributionShow details

Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lithuania, Malaysia, Mexico, Moldova, Morocco, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Tunisia, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Venezuela and Vietnam.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2624-2026

Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…

  • Pulmonary Valve Prosthesis Percutaneously Delivered
  • Process control
Medtronic Heart Valves…CAJul 8, 2026
Class IOngoingZ-2622-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026
Class IOngoingZ-2621-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026