Class IIOngoingZ-0924-2025

Intuitive Surgical, Inc. recalls 8MM

Intuitive Surgical, Inc.Sunnyvale, CA, United StatesReported Jan 22, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Due to an increase in complaints for pitch cable failures related to forceps and small Graptor instruments

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • 8MM, Small Graptor REF 470318 Is a grasping retractor Instrument used with the da Vinci X/Xi Surgical Systems in a wide variety of procedures to perform a variety of surgical tasks including manipulation and retraction of tissue
    UDI-DI code
    359 affected lots
    S10131209S10131210470318-07S10140617S10140618S10140619S10140620S10140717
    +351 moreS10140801S10140812S10140821S10140823S10140915S10141008S10141120S10141205S10141219S10150108S10150120S10150122S10150206S10150212S11140620470318-08N10151230N10160104N10160113N10160203N10160218N10160219N10160307N10160317N10160330N10160404N10160406N10160420N10160426N10160525N10160607N10160616N10160725N10160726N10160802N10160810N10160824N10160901N10160914N10160927N10161010N10161020N10161102N10161111N10161128N10161209N10170119N10170203N10170222N10170309N10170330N10170419N10170502N10170512N10170518N10170612N10170623N10170703N10170725N10170728N10170801N10170802N10170822N10170823N10170906N10170914N10171012N10171101N10171115N10171124N10171206N10171219N10171227N10180122N10180131N10180209N10180221N10180228N10180312N10180328N10180410N10180413N10180511N10180518N10180606N10180612N10180613N11170725N11170728N12170725S10141211S10150223S10150312S10150326S10150408S10150417S10150430S10150506S10150611S10150706S10150717S10150801S10150827S10150914S10151007S10151012S10151028S10151104S10151215S10160125S10160217S10160413S10160512S10160614S10160622S10160715S10160719S10160913S10161221S10170525S10170710S11150511S11150603S11150624S11150728S11150813S11150904S11150926S11151023S11151119S11151201S11151231S12150819S12150918S12151001S12151008S12151019S12151219S13150820T10170822T10170929T10171004T10171030T10171127470318-10K10210921K10210927K10211004K10211011K10211018K10211103K10211108K10211116K10211117K10211122K10211206K10211213K10220217K10220221K10220228K10220307K10220314K10220404K10220425K10220511K10220516K10220613K10220725K10220802K10220808K10220815K10220912K10220926K10221016K10221030K10221102K10221106K10221113K10221117K10221118K10230212K10230219K10230226K10230420K10230504K10230511K10230518K10230525K10230601K10230608K10230615K10230706K10230713K10230803K10230817K10230831K10230907K10230914K10230921K10230928K10231012K10231019K10231027K10231130K10231207K10240104K10240116K10240125K10240222K10240327K10240404K10240516K10240523K10240606K10240711K10240725K10240822K10240829K10240905K11211103K11220110K11220228K11220511K11220516K11220912K11220926K11221030K11230608K12220511K12221102N10180620N10180704N10180711N10180719N10180823N10180828N10180831N10180920N10181001N10181017N10181022N10181102N10181106N10181113N10181115N10181123N10181128N10181203N10181206N10181210N10190103N10190107N10190129N10190206N10190211N10190216N10190222N10190223N10190226N10190304N10190307N10190310N10190322N10190330N10190407N10190411N10190425N10190507N10190515N10190519N10190521N10190531N10190602N10190701N10190715N10190722N10190729N10190812N10190819N10190826N10190902N10190909N10190917N10190923N10190930N10191007N10191028N10191104N10191125N10191202N10200106N10200211N10200302N10200309N10200317N10200323N10200330N10200406N10200413N10200420N10200427N10200504N10200629N10200713N10200727N10200803N10200810N10200817N10200928N10201012N10201019N10201027N10201110N10201117N10201123N10201130N10201207N10201214N10210104N10210111N10210118N10210125N10210202N10210208N10210215N10210222N10210301N10210308N10210316N10210322N10210405N10210412N10210503N10210518N10210531N10210607N10210621N10210705N10210719N10210831N10210906N10210913N11200323N11200727N11210503N13210831U10210805U10220209U10220301U10220328U10220506

What the firm is doing

On 12/19/2024, the firm sent an "Field Safety Notice Urgent Medical Device Notification" via FedEx (US), email/hand-delivery (OUS), informing customers that Intuitive Surgical has observed an increase in complaints regarding pitch cable failure on the Tenaculum Forceps and Small Graptor retractor devices. A failure of the pitch cable can be fraying or complete break. Customers are instructed: 1. As a reminder, when using the Tenaculum Forceps and the Small Graptor instruments, refer to and follow the instructions, warnings and cautions provided in the General Overview and EndoWrist Instrument chapters of the da Vinci X/Xi Instruments and Accessories User Manual and Reprocessing Instructions User Manual. 2. Display this communication with affected systems; ensures it is in a place likely to be seen/viewed by operators. 3. Complete the Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 4. Retain a copy of this letter and the acknowledgement form for your files. 5. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject devices via the standard complaint process. For questions/information/further support - contact Clinical Sales Representative or Customer Service: "North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com. "Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com "South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com "Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com "India: +1-800-103-6952 (9 am to 6 pm IT) "Taiwan: +0800-86-8181 (9 am to 6 pm CT)

DistributionShow details

U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WV, WI, and WY. The countries of Canada, Austria, Belgium, Denmark, Finland, France, French Guiana, Germany, Guadeloupe, Ireland, Luxembourg, Martinique, Monaco, Netherlands, Norway, Reunion, Sweden, Switzerland, United Kingdom, India, Japan, South Korea, Taiwan, Brazil, Argentina, Australia, Bulgaria, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Ecuador, Greece, Hong Kong, Hungary, Iceland, Israel, Italy, Kuwait, Lebanon, Malaysia, Malta, Mexico, New Zealand, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Thailand, Turkey, United Arab Emirates, Venezuela, Vietnam.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.