Medline Industries, LP recalls MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed
Reason for recall
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L
Lot / code information
UDI/DI 40080196294344, All lots
What the firm is doing
Medline issued an URGENT MEDICAL DEVICE RECALL notice with an enclosed URGENT MEDICAL DEVICE FIELD CORRECTION NOTICE to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 orrecalls@medline.com.
DistributionShow detailsHide
US, CA, PA, GUAM, VI
A reasonable probability that use of the product will cause serious adverse health consequences or death.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1015-2026
- FDA device classification · LLIOfficial FDA classification for this device type
- CFR regulation · 21 CFR 880.5100The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medline Industries, LPSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class I recalls
Draeger, Inc. recalls VentStar Resus Neo. Neonatal. 25 Units per Box. Part Number: MP00310 Disposable brea…
Potential cracks in the hose, which could lead to leakage or the hose falling off.
- Positive End Expiratory Pressure Breathing Attachment
- Under Investigation by firm
Medline Industries, LP recalls Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire.…
- Manual Bed
- Device Design
Medline Industries, LP recalls Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire.…
- Adjustable Home-Use Ac-Powered Therapeutic Bed
- Device Design