Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-1085-2026

Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS

Medline Industries, LPNorthfield, IL, United StatesReported Jan 21, 2026 · 5 months ago
Legal News Analyst ·

Reason for recall

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359

Lot / code information

Lot #
25EBD732; 2) UDI/DI 10198459326318 (EA) 40198459326319 (CS)
Lot #
25FBA103

What the firm is doing

Medline Industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/16/2025 via email and USPS first class mail. The notice explained the issue, potential risk, and requested the following: "1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantityof affected product you have in inventory on the form. Even if you do not have any affected productin inventory, please complete and submit the form. Website link: https://recalls.medline.comRecall Reference #: R-25-145-FGX5 Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4.If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com."

DistributionShow details

US Nationwide distribution.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.