ICU Medical Inc. recalls Portex Spinal Tray

ICU Medical Inc.Minneapolis, MN, United StatesReported Jan 21, 2026 · 5 months ago
Legal News Analyst ·

Reason for recall

Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Portex Spinal Tray, Item No. A3729-24A3595
    UDI-DI 30351688075691
    5 affected lots
    61462041/22/26A3595350195171082116157980

What the firm is doing

On December 19. 2025, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were informed that affected lots may have a crack present on the syringe collar. Customers were instructed to discontinue use of affected lots, and to destroy products following their institution's process for destruction. If destruction is not immediately possible, affected product should be quarantined until disposal.

DistributionShow details

US Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO, NE, NJ, NY, OH, OK, TX, WA.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.