Class IOngoingZ-1159-2025

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M12

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Best, NetherlandsReported Feb 26, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Possibility of the patient falling from the table related to the mattress used on the patient table.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Azurion 7 M12; Catalog numbers: (1) 722078, (2) 722223, (3) 722233 (OUS ONLY).

Lot / code information

Catalog #
(1) 722078, (2) 722223, (3) 722233 (OUS ONLY)
UDI
(1) 00884838085251, (2) 00884838099241, (3) 00884838116764
Serial #
(1) 465, 437, 438, 143, 302, 559 — +1702 moreShow all
(1) 465, 437, 438, 143, 302, 559, 170, 824, 626, 697, 278, 491, 215, 720, 612, 788, 141, 644, 279, 402, 290, 524, 415, 890, 412, 166, 136, 539, 532, 538, 291, 483, 497, 576, 500, 379, 753, 677, 709, 649, 441, 305, 289, 22, 257, 258, 187, 803, 399, 613, 121, 396, 536, 537, 488, 566, 833, 456, 478, 479, 224, 309, 719, 163, 921, 296, 364, 684, 239, 383, 249, 125, 829, 513, 376, 627, 725, 728, 611, 308, 522, 862, 1007, 661, 298, 571, 654, 656, 657, 658, 292, 844, 876, 480, 481, 410, 271, 493, 327, 550, 723, 269, 185, 902, 445, 360, 232, 425, 635, 546, 748, 294, 906, 603, 645, 870, 544, 545, 523, 520, 521, 556, 557, 558, 293, 840, 598, 384, 317, 535, 359, 700, 251, 786, 240, 256, 436, 585, 31, 165, 247, 733, 204, 205, 206, 888, 923, 195, 197, 162, 517, 218, 797, 444, 489, 222, 361, 362, 977, 30, 579, 223, 584, 618, 909, 193, 280, 962, 893, 287, 263, 174, 156, 27, 194, 953, 356, 182, 958, 847, 959, 713, 610, 325, 284, 527, 189, 614, 600, 864, 760, 490, 103, 140, 184, 506, 342, 673, 674, 675, 676, 23, 112, 750, 199, 363, 217, 151, 152, 160, 815, 695, 135, 985, 1058, 1077, 1047, 940, 942, 945, 946, 947, 948, 1008, 1032, 749, 918, 919, 989, 984, 1011, 701, 907, 1015, 735, 736, 856, 898, 1037, 1051, 1086, 1000, 987, 819, 820, 1057, 1081, 798, 992, 1036, 993, 986, 710, 1069, 1070, 1071, 1072, 971, 849, 1004, 1068, 1043, 1059, 1034, 1046, 1064, 1065, 1066, 1067, 1009, 716, 717, 994, 1075, 1074, 1084, 1085, 1073, 1079, 1062, 1063, 926, 687, 688, 1056, 1045, 1048, 1049, 1033, 1035, 899, 1061, 970, 875, 1026, 964, 965, 1039, 1080, 1052, 407, 408, 505, 581, 850, 643, 314, 316, 976, 806, 502, 200, 943, 957, 1013, 1025, 354, 991, 551, 775, 776, 320, 295, 547, 123, 319, 525, 318, 586, 879, 127, 128, 130, 132, 133, 686, 805, 346, 347, 398, 24, 564, 604, 690, 769, 570, 572, 565, 470, 562, 378, 414, 727, 38, 703, 865, 621, 348, 881, 453, 468, 142, 374, 375, 126, 726, 751, 469, 114, 689, 845, 430, 149, 499, 625, 897, 261, 472, 254, 255, 267, 268, 368, 369, 243, 244, 731, 202, 282, 283, 351, 442, 636, 40, 172, 642, 477, 134, 262, 341, 563, 486, 633, 21, 705, 286, 336, 512, 158, 357, 233, 397, 896, 281, 52, 190, 366, 617, 496, 461, 463, 777, 129, 270, 528, 529, 638, 431, 712, 714, 219, 203, 575, 345, 934, 498, 574, 577, 853, 668, 779, 548, 37, 620, 1031, 306, 377, 381, 988, 782, 237, 540, 601, 424, 596, 597, 662, 440, 628, 952, 422, 259, 935, 510, 33, 954, 196, 371, 10, 148, 476, 507, 762, 503, 950, 854, 846, 213, 227, 561, 663, 380, 778, 462, 464, 891, 542, 968, 680, 315, 831, 841, 905, 830, 843, 842, 682, 501, 323, 683, 567, 884, 602, 153, 154, 173, 418, 761, 139, 178, 179, 264, 265, 266, 511, 526, 746, 201, 353, 487, 495, 186, 122, 338, 518, 303, 304, 509, 515, 432, 433, 310, 18, 582, 19, 416, 530, 664, 11, 155, 435, 715, 685, 508, 543, 516, 16, 908, 167, 446, 12, 225, 913, 901, 198, 504, 961, 457, 967, 647, 367, 980, 981, 982, 328, 343, 344, 764, 770, 771, 772, 773, 972, 800, 648, 191, 652, 917, 578, 912, 169, 426, 650, 471, 419, 236, 979, 938, 667, 619, 541, 738, 324, 365, 1003, 835, 836, 929, 975, 252, 36, 242, 622, 623, 629, 207, 208, 209, 210, 211, 300, 301, 795, 32, 241, 785, 49, 50, 51, 145, 250, 774, 474, 475, 452, 624, 447, 839, 333, 837, 404, 681, 26, 692, 183, 651, 177, 615, 423, 554, 181, 816, 616, 696, 672, 175, 855, 434, 358, 722, 439, 789, 552, 1010, 871, 874, 1029, 1030, 869, 939, 180, 990, 28, 29, 659, 339, 340, 20, 458, 421, 138, 157, 752, 355, 307, 553, 593, 944, 857, 382, 428, 599, 467, 630, 729, 102, 370, 450, 147, 429, 449, 757, 216, 743, 759, 334, 451, 960, 159, 400, 401, 411, 124, 13, 14, 15, 818, 459, 460, 352, 46, 39, 311, 312, 277, 718, 698, 150, 693, 694, 589, 784, 482, 235, 792, 590, 485, 796, 838, 349, 1005, 226, 920, 272, 877, 322, 978, 455, 641, 631, 632, 220, 221, 146, 609, 448, 466, 704, 801, 802, 492, 533, 41, 42, 43, 44, 45, 669, 35, 321, 332, 331, 817, 699, 144, 594, 192, 799, 608, 878, 732, 768, 637, 670, 335, 253, 549, 892; (2) 580, 669, 282, 357, 367, 164, 799, 124, 1060, 1107, 36, 570, 665, 455, 457, 301, 456, 704, 131, 441, 277, 616, 750, 872, 877, 816, 1144, 365, 390, 1015, 69, 1103, 599, 601, 602, 470, 171, 432, 352, 1084, 340, 892, 597, 97, 242, 361, 161, 205, 627, 514, 628, 198, 371, 229, 230, 262, 261, 328, 727, 560, 561, 463, 487, 1113, 214, 318, 458, 883, 889, 508, 566, 256, 351, 596, 1140, 637, 638, 433, 728, 937, 944, 945, 951, 533, 323, 180, 659, 377, 435, 439, 1051, 1008, 452, 453, 681, 77, 127, 1179, 47, 67, 932, 252, 626, 126, 905, 386, 565, 308, 449, 886, 887, 1150, 577, 85, 312, 324, 332, 585, 96, 44, 343, 1231, 221, 880, 746, 966, 105, 572, 852, 801, 269, 192, 197, 87, 204, 1181, 1114, 1161, 188, 130, 52, 379, 440, 668, 888, 177, 223, 503, 196, 882, 208, 1123, 104, 334, 388, 581, 855, 207, 189, 479, 329, 448, 495, 571, 211, 186, 187, 631, 567, 1001, 1097, 829, 206, 210, 336, 373, 258, 679, 1088, 125, 1065, 391, 454, 37, 119, 136, 370, 116, 154, 349, 387, 89, 133, 99, 100, 912, 697, 698, 139, 355, 871, 1236, 1035, 1036, 775, 903, 598, 867, 619, 445, 222, 854, 1191, 756, 436, 437, 438, 325, 999, 879, 787, 1033, 530, 579, 553, 278, 779, 78, 1129, 381, 926, 706, 1165, 202, 38, 859, 430, 940, 313, 314, 201, 185, 1083, 857, 860, 382, 1109, 846, 847, 169, 102, 649, 645, 646, 647, 648, 1143, 108, 786, 578, 650, 1128, 986, 1057, 990, 991, 987, 1055, 1169, 780, 781, 825, 1245, 758, 1171, 652, 900, 651, 604, 843, 782, 1124, 1172, 957, 1176, 901, 763, 764, 765, 767, 1013, 1014, 996, 997, 838, 1026, 798, 823, 979, 686, 687, 688, 689, 1275, 692, 953, 611, 644, 1218, 1016, 1115, 797, 917, 1225, 762, 1214, 517, 1198, 634, 678, 924, 443, 132, 296, 297, 41, 914, 461, 30, 808, 298, 552, 165, 24, 117, 195, 1160, 793, 794, 795, 173, 212, 267, 268, 306, 243, 655, 656, 662, 664, 537, 694, 419, 666, 413, 271, 468, 270, 255, 573, 1100, 558, 559, 563, 564, 33, 541, 224, 606, 965, 354, 711, 356, 934, 414, 416, 417, 671, 403, 933, 804, 710, 287, 114, 311, 863, 864, 915, 925, 237, 26, 919, 238, 239, 574, 289, 286, 288, 290, 1066, 247, 961, 963, 906, 824, 249, 477, 814, 155, 8
Serial #
Added 5/2/2025**** ; 1038, 1052, 1061, 1067, 1106, 1130, 1132, 1185, 1188, 1211, 1241, 1242; (3) OUS ONLY 6, 7, 8, 9, 24 ****
Serial #
Added 5/2/2025**** 5, 11, 12, 15, 16, 17 — +42 moreShow all
Added 5/2/2025**** 5, 11, 12, 15, 16, 17, 18, 19, 20, 48, 27, 28, 29, 31, 32, 35, 37, 38, 39, 40, 43, 44, 46, 47, 48, 50, 51, 55, 57, 58, 60, 61, 62, 71, 74, 77, 80, 81, 85, 86, 87, 89, 90, 91, 97, 102, 104, 134

What the firm is doing

On January 14, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 4. Actions that should be taken by the customer / user in order to prevent risks for patients or users " Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. " Follow the additional instructions on the use of the mattress included in Appendix B. These instructions include information for the transfer of the patient and the correct use and positioning of the mattress. The information of this Appendix is being included in an Addendum to the Instructions for Use. Additionally, Philips is creating a Quick Reference Card. The Addendum and the Quick Reference Card will be available as from end of March 2025. Please refer to Appendix C on how to download the Addendum to the Instructions for Use and the Quick Reference Card. " In case that the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through the email account: IGT_Recalls@philips.com " Please complete and return the Response form attached to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issue and required actions to be taken. " Should you experience the issue reported in this letter, please report the event to Philips through your local Philips representative (1- 800-722-9377). 5. Actions planned by Philips Image Guided Therapy Systems to correct the problem Philips is informing customers of this issue through this Urgent Medical Device Correction. Additionally, Philips is releasing an Addendum to the Instructions for Use and a Quick Reference Card that will be available for customers at the end of March 2025. This Urgent Medical Device Correction letter has been reported to the appropriate Regulatory Agencies.

DistributionShow details

Worldiwde distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, JAMAICA, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2624-2026

Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…

  • Pulmonary Valve Prosthesis Percutaneously Delivered
  • Process control
Medtronic Heart Valves…CAJul 8, 2026
Class IOngoingZ-2622-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026
Class IOngoingZ-2621-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026