Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: 1) PACER / CATH LAB PACK-LF
Reason for recall
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Medline medical procedure convenience kits labeled as: 1) PACER / CATH LAB PACK-LF, Kit SKU DYNJ0972604A; 2) PACEMAKER/ICD PACK, Kit SKU DYNJ33822L; 3) VENOUS ACCESS PACK, Kit SKU DYNJ45640B; 4) PACEMAKER SET UP, Kit SKU DYNJ63377D; 5) PACK PACEMAKER CUSTOM,9909178, Kit SKU DYNJ65011D; 6) HEART BASIC OPEN PACK, Kit SKU DYNJ69143C; 7) SCC VEIN PACK, Kit SKU DYNJ69313; 8) CV PACK, Kit SKU DYNJ86007B.
Lot / code information
- Lot #
- 25KMD144; Medline Kit SKU DYNJ33822L: UDI/DI 10198459385582 (EA) 40198459385583 (CS)
- Lot #
- 25KMD419; Medline Kit SKU DYNJ45640B: UDI/DI 10193489494983 (EA) 40193489494984 (CS)
- Lot #
- 25KMD793; Medline Kit SKU DYNJ63377D: UDI/DI 10198459514579 (EA) 40198459514570 (CS)
- Lot #
- 25KMI988; Medline Kit SKU DYNJ65011D: UDI/DI 10198459017018 (EA) 40198459017019 (CS)
- Lot #
- 25KMH161; Medline Kit SKU DYNJ69143C: UDI/DI 10198459478505 (EA) 40198459478506 (CS)
- Lot #
- 25LMB211; Medline Kit SKU DYNJ69313: UDI/DI 10193489931822 (EA) 40193489931823 (CS)
- Lot #
- 25KMJ489; Medline Kit SKU DYNJ86007B: UDI/DI 10198459260452 (EA) 40198459260453 (CS)
- Lot #
- 25KDB551
What the firm is doing
Medline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-25-251-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
DistributionShow detailsHide
US Nationwide distribution.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1188-2026
- FDA device classification · OEZOfficial FDA classification for this device type
- CFR regulation · 21 CFR 870.1200The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medline Industries, LPSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm
