CareFusion 303, Inc. recalls BD Pyxis: MedFlex
Reason for recall
Automated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential delays in retrieving contents/ accessing medications that may result from power loss or potential automated dispensing device malfunctions.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
BD Pyxis: MedFlex1 code
- 1119-00
MedBank Tower1 code
- 1145-00
MedBank MedPass1 code
- 139088-01 and associated BD Pyxis user/software/safety guides and quick reference guides.
Lot / code information
- Part / Item #
155018-03) ii. BD Pyxis MedBank Cabinet QRG (PN 155023-03) iii. BD Pyxis MedBank RxNow CR Station QRG (PN 155026-03) iv. BD Pyxis MedBank RxNow Solution QRG 2.0 (PN 155027-02) v. BD Pyxis MedBank MedP…Show all
155018-03) ii. BD Pyxis MedBank Cabinet QRG (PN 155023-03) iii. BD Pyxis MedBank RxNow CR Station QRG (PN 155026-03) iv. BD Pyxis MedBank RxNow Solution QRG 2.0 (PN 155027-02) v. BD Pyxis MedBank MedPass QRG (PN 155022-03) vi. BD Pyxis MedBank QFill Station QRG (PN 155020-02) vii. BD Pyxis MedBank myQLink Cloud Software QRG 3.0 (PN 155021-02) REF/UDI-DI/- Serial #
1119-00/10885403512704/13517006, 13523350, 13754724, 13767953, 13789575, 13801727 — +1512 moreShow all
1119-00/10885403512704/13517006, 13523350, 13754724, 13767953, 13789575, 13801727, 13826187, 13879544, 13883357, 13884848, 13934472, 13939327, 13948724, 14032802, 14037891, 14158187, 14159916, 14168580, 14172832, 14175162, 14182716, 14202337, 14220872, 14222746, 14228149, 14228150, 14244101, 14254813, 14254816, 14254812, 14254815, 14275112, 14278095, 14317945, 14317946, 14324386, 14332058, 14347747, 14347748, 14349106, 14348708, 14348709, 14352459, 14365786, 14365785, 14374965, 14374963, 14374964, 14377079, 14378428, 14377908, 14407523, 14407524, 14420205, 14420213, 14420211, 14420207, 14420208, 14420206, 14420223, 14420222, 14420212, 14441382, 14442352, 14442359, 14442686, 14442685, 14442585, 14442586, 14445911, 14456782, 14456757, 14458260, 14458256, 14458259, 14472490, 14472338, 14472980, 14472990, 14472975, 14472981, 14472970, 14472977, 14472982, 14472971, 14472987, 14472973, 14472983, 14472972, 14472961, 14472985, 14472974, 14472989, 14472966, 14472988, 14472984, 14472964, 14472962, 14472968, 14472969, 14472967, 14472979, 14475836, 14475827, 14475839, 14475832, 14475831, 14475830, 14475828, 14475829, 14475837, 14475834, 14475835, 14532831, 14533089, 14540092, 14540091, 14540107, 14540097, 14540103, 14545019, 14549918, 14549926, 14549535, 14549924, 14549537, 14549913, 14549910, 14549934, 14549928, 14549543, 14549534, 14549542, 14549546, 14579394, 14602051, 14602043, 14602044, 14602049, 14602048, 14602050, 14618948, 14635117, 14635121, 14635120, 14635118, 14635122, 14649376, 14651893, 14656794, 14672001, 14678266, 14678259, 14678251, 14678262, 14678240, 14678261, 14678246, 14678258, 14678256, 14695735, 14695732, 14695737, 14695731, 14697048, 14697051, 14697049, 14709300, 14711347, 14711353, 14742553, 14742644, 14743937, 14743923, 14743935, 14743929, 14743932, 14743938, 14743925, 14743933, 14743934, 14743931, 14787943, 14787944, 14787922, 14787923, 14787942, 14787939, 14787919, 14787940, 14803084, 14834314, 14861749, 14861747, 14868948, 14883876, 14888627, 14888639, 14894201, 14894199, 14894200, 14894202, 14897578, 14897577, 14925753, 14942508, 14951354, 14951344, 14951377, 14951386, 14951338, 14958134, 14977993, 14990472, 14990476, 15024805, 15070091, 15086034, 15086026, 15125781, 15125782, 15125773, 15125779, 15125784, 15125777, 15125778, 15125776, 15125772, 15169037, 15176582, 15176580, 15201877, 15201878, 15206270, 15206267, 15206269, 15210095, 15210091, 15210094, 15210089, 15210104, 15210088, 15210090, 15210106, 15213239, 15216160, 15223570, 15223560, 15223568, 15223571, 15229115, 15229108, 15229107, 15229119, 15229118, 15236627, 15240793, 15242603, 15242599, 15242596, 15242602, 15242605, 15242606, 15245775, 15245774, 15245776, 15245777, 15253442, 15254778, 15254776, 15254774, 15254772, 15254775, 15268840, 15268836, 15268851, 15268838, 15268856, 15268850, 15268835, 15279442, 15279443, 15279444, 15279455, 15279458, 15279454, 15290675, 15294474, 15299940, 15308354, 15317099, 15317097, 15317100, 15317104, 15317101, 15317102, 15322148, 15322145, 15340242, 15340240, 15340245, 15340241, 15340239, 15340251, 15340243, 15340238, 15356973, 15356977, 15361554, 15361559, 15361555, 15361553, 15369939, 15389013, 15389673, 15391644, 15391311, 15392443, 15392451, 15392447, 15392449, 15393558, 15396726, 15416741, 15460705, 15460996, 15460706, 15460997, 15460707, 15461858, 15463097, 15463095, 15463096, 15464898, 15470033, 15470035, 15470034, 15471183, 15473693, 15473692, 15474596, 15474551, 15492248, 15492211, 15498908, 15500854, 15509359, 15509361, 15509364, 15509365, 15509366, 15513835, 15513825, 15513834, 15513826, 15513833, 15513830, 15513831, 15517161, 15519704, 15519700, 15521199, 15521203, 15521200, 15521201, 15554707, 15557611, 15561697, 15564895, 15564897, 15564894, 15564893, 15564896, 15583141, 15584199, 15613096, 15614585, 15622613, 15622611, 15622635, 15638496, 15638495, 15643124, 15643125, 15643128, 15647364, 15647367, 15652143, 15652147, 15652138, 15652145, 15661682, 15661681, 15679247, 15679269, 15679254, 15679243, 15679263, 15679237, 15679245, 15682324, 15691543, 15736824, 15736832, 15736823, 15748541, 15748544, 15748543, 15779891, 15788872, 15788870, 15788868, 15803799, 15835058, 15835059, 15853794, 15854592, 15854989, 15855085, 15855653, 15855651, 15855650, 15855649, 15855652, 15906182, 15923716, 15924566, 15924567, 15929989, 15931343, 15933057, 15933056, 15937957, 15940593, 15940620, 15942158, 15942157, 15942159, 15956999, 15957002, 16030464, 16039729, 16045065, 16053361, 16063826, 16066740, 16081624, 16084502, 16089562, 16089561, 16089912, 16090406, 16090401, 16090325, 16090399, 16090308, 16090687, 16091227, 16091885, 16112808, 16112853; 1145-00/10885403517167/15383483, 15389562, 15422138, 15422139, 15442792, 15500700, 15500693, 15504839, 15504841, 15518256, 15518261, 15550844, 15550852, 15550857, 15569626, 15598106, 15600761, 15639702, 15654701, 15654698, 15662824, 15662821, 15684799, 15684812, 15715761, 15715760, 15715748, 15715742, 15723806, 15723817 15728652, 15728657, 15747087, 15761382, 15770179, 15770181, 15770187, 15770221, 15770215, 15770217, 15770207, 15770172, 15806707, 15806718, 15867348, 15928765, 15930653, 15932837, 15935612, 15941687, 15957628, 15958445, 15958444, 15962003, 15962008, 15962006, 15962002, 15962016, 15961996, 15961979, 15961986, 15962018, 15967374, 15967375, 15967376, 15978031, 16006031, 16295928, 16295929, 16295930, 16295931, 16296379, 16296380, 16296360, 16297667, 16297668, 16297669, 16297671, 16300643, 16300642, 16300639, 16300641, 16300640, 16301023, 16301024, 16301018, 16301021, 16301020, 16301015, 16301016, 16302339, 16302328, 16302327, 16302337, 16302343, 16302338, 16302331, 16302332, 16302333, 16302342, 16302340, 16302329, 16302330, 16303465, 16303469, 16303463, 16303466, 16303468, 16303467, 16305129, 16305136, 16305134, 16305130, 16305133, 16306411, 16306401, 16306402, 16306413, 16306408, 16306419, 16306403, 16306397, 16306399, 16306404, 16306393, 16306395, 16308990, 16308989, 16308977, 16308973, 16308987, 16308979, 16308993, 16308998, 1630 …
What the firm is doing
On 1/10/2025, correction notices were mailed or emailed to Directors of Nursing or Nurse Managers, Directors of Pharmacy or Pharmacist In Charge, and Facility Administrators informing them to take the following actions: Customers should do the following to prevent delays in access: a. Ensure staff are aware of updates to User Guides, including: automated dispensing devices should not be used as a crash cart or code box. Medication or items may not be immediately accessible for an emergent situation causing a delay in treatment which could lead to serious injury or death. b. Have device keys available on-site in a secure location known to designated staff c. Train staff on how to manually access the device d. Ensure downtime policies and procedures are implemented and staff are appropriately trained e. Enable Over-ride mode permissions For pharmacies providing automated dispensing device products to clinical users that are not physically located at the pharmacy site, please ensure clinical users are aware of the risk of delay in access and provide access to keys. Firm will be strengthening labeling regarding delays in dispensing medication. Report any issues associated with the above devices via phone at 866-930-9251. Post these notices with the device until labeling has been updated. Complete and return the response form via email to BDRC54@bd.com If you require further assistance, please contact the firm: 1) Technical Support: Website: bd.com/self-service, Phone: 1-866-930-9251, Hours: 24 hours, 7 days a week. 2) BD Recall Operations, BDRC54@bd.com, Fax: (312) 949-0081. 3) Customer Letter Support, Email: PyxisQuestions@BD.com
DistributionShow detailsHide
US Nationwide distribution including in the states of AZ, FL, NJ, KY, GA, CA, IL, MD, WI, OH, NC, MS, NY, MO, PA, OR, IN, TX, UT, OK, MI, WA, AL, NV, ME, WV, NE, DC, MA, MT, KS, CO, TN, IA, MN, AR, ID, SC, SD, CT, HI, VA, ND, AK, NM, VT, LA, WY.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1206-2025
- FDA device classification · BRYOfficial FDA classification for this device type
- CFR regulation · 21 CFR 868.6100The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find CareFusion 303, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm