CareFusion 303, Inc. recalls BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a mo…
Reason for recall
Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administrationUDI-DI codesCatalog # Number
What the firm is doing
On 01/23/2025, the firm sent via FedEx, 2-day mail and email an "URGENT MEDICAL DEVICE LABELING CORRECTION" Letter to customers informing them that Through an internal review, BD has identified that in the BD Alaris" System User Manual v 12.3.1 (p/n #P00000455), 12.3.2 (p/n# P00000827) and 12.4 (p/n# P00000828), there is an incorrect note on the Pump Module Set Loading instructions. The manuals incorrectly instruct users to leave the safety clamp in the open position when reloading an infusion set. The correct procedure is to leave the safety clamp in the closed position. Following the incorrect note could potentially disable a safety feature that could result in an accidental over-infusion if the roller clamp is left open. Customers are instructed to: 1. Post the Customer Notification on or near the affected systems as a reminder of the incorrect labeling issue until the Manual/Directions for Use (DFU) have been corrected. 2. Complete the attached Customer Response Form and return to the BD contact noted on the form within 14 days, so that BD may acknowledge your receipt of this notification per FDA requirements and prioritize remediation scheduling. Actions BD is taking: 1. BD is in the process of updating the impacted user manuals. 2. BD will send customers follow-up notifications once the update of user manuals is completed. BD expects an updated user manual to be available by April 2025. For Questions or assistance: Contact - North American Regional Complaint Center for Product Complaints and Technical Questions at: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted. Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com Contact Post Market Quality for Recall Questions at: Email : BDRC39@bd.com Fax No.: 312-949-0576
DistributionShow detailsHide
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and Guantanamo Bay, Cuba O.U.S.: Canada
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1231-2025
- FDA 510(k) clearance · K211218The device's official FDA premarket clearance record
- FDA device classification · FRNOfficial FDA classification for this device type
- CFR regulation · 21 CFR 880.5725The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find CareFusion 303, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm