Class IIOngoingZ-1238-2026

Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6…

Medtronic MiniMed, Inc.Northridge, CA, United StatesReported Feb 11, 2026 · 5 months ago
Legal News Analyst ·

Reason for recall

A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more. The MiniMed 780G system is intended for continuous delivery of basal insulin at selectable rates, and the administration of insulin boluses at selectable amounts for the management of type 1 diabetes mellitus in persons seven years of age and older requiring insulin. The system is also intended to continuously monitor glucose values in the fluid under the skin. The MiniMed 780G system includes SmartGuard technology, which can be programmed to automatically adjust insulin delivery based on continuous glucose monitoring (CGM) sensor glucose values and can suspend delivery of insulin when the SG value falls below or is predicted to fall below predefined threshold values
    UDI-DI codes
    400 affected lots
    NG3666779HNG3811803HNG3926926HNG3944480HNG4116699HNG4173817HNG3334724HNG3750453H
    +392 moreNG3874105HNG3924178HNG3926632HNG3927673HNG3944520HNG4023801HNG4029254HNG4097850HNG4146571HNG4170515HNG4200310HNG4177489HNG4178626HNG2752574HNG3423286HNG3838934HNG3926282HNG3948645HNG4017463HNG4010052HNG4174745HNG4178395HNG3502087HNG3634564HNG3756694HNG3871741HNG3919620HNG3927825HNG4164005HNG3729206HNG3839883HNG3884748HNG3908444HNG3961353HNG4068853HNG4179772HNG3106884HNG3116129HNG3271492HNG3351525HNG3458389HNG3464645HNG3490545HNG3482017HNG3520991HNG3593959HNG3515336HNG3467707HNG3563980HNG3654085HNG3653614HNG3662418HNG3631555HNG3686291HNG3685575HNG3721941HNG3722374HNG3751345HNG3762687HNG3773707HNG3804837HNG3785876HNG3815941HNG3809816HNG3809672HNG3809900HNG3811760HNG3857851HNG3848444HNG3865466HNG3871979HNG3923251HNG3928992HNG3946093HNG3944898HNG3945847HNG3966122HNG3925413HNG3984252HNG4071539HNG3891538HNG4099361HNG4097100HNG3966344HNG4115711HNG4170457HNG4170650HNG4171411HNG4179308HNG4165581HNG4166028HNG4174665HNG4176618HNG4197135HNG4198605HNG4194835HNG4213027HNG2811465HNG2922632HNG3050369HNG2473989HNG3110782HNG3165546HNG3245352HNG3491506HNG3504053HNG3503998HNG3467406HNG3727425HNG3718393HNG3741705HNG3740402HNG3754549HNG3743216HNG3838750HNG3859322HNG3846048HNG3851889HNG3869122HNG3878439HNG3920428HNG3941436HNG3919322HNG3927745HNG3946295HNG3923212HNG3941150HNG3978427HNG3945129HNG3952217HNG4041650HNG4019446HNG4010956HNG4025469HNG4079778HNG4089972HNG4087937HNG4082944HNG4068836HNG4075347HNG4078121HNG4088929HNG4171211HNG4177285HNG4174489HNG4177168HNG4176329HNG4208285HNG3666827HNG3695566HNG3713071HNG3717692HNG3745851HNG3749951HNG3735248HNG3774228HNG3848457HNG3848901HNG3849344HNG3855096HNG3874209HNG3891021HNG3878609HNG3872267HNG3892797HNG3925816HNG3944275HNG3938415HNG3934980HNG3978206HNG3997696HNG4023617HNG4067807HNG4068186HNG4068276HNG4097308HNG4086943HNG4127318HNG4116303HNG4168668HNG4170561HNG4177312HNG4187583HNG4200314HNG4188462HNG4194675HNG4196463HNG4193920HNG2983018HNG3226459HNG3253416HNG3375428HNG3389508HNG3455845HNG3290714HNG3544255HNG3496998HNG3664014HNG3665167HNG3741756HNG3781060HNG3851847HNG3874189HNG3873296HNG3904518HNG3923064HNG3920073HNG3941074HNG3978585HNG3946268HNG3963748HNG3950286HNG3998081HNG4068503HNG4066883HNG3890599HNG4102568HNG4104044HNG4117340HNG3998278HNG4127913HNG4171084HNG4174122HNG4174291HNG4182715HNG4186723HNG4187005HNG4176580HNG4195536HNG3458407HNG3473827HNG3477699HNG3613124HNG3651486HNG3727632HNG3732603HNG3732097HNG3742868HNG3762837HNG3761312HNG3813562HNG3822747HNG3857855HNG3922024HNG3920283HNG3924778HNG3951674HNG3940453HNG3985400HNG4018435HNG4060209HNG3900942HNG4102555HNG4124451HNG4174561HNG4174265HNG4163176HNG4182361HNG4164206HNG4186544HNG4208987HNG4209038HNG4209772HNG3163771HNG3411122HNG3458356HNG3491051HNG3567668HNG3544931HNG3713706HNG3718296HNG3809723HNG3818145HNG3809695HNG3864681HNG3887795HNG3679742HNG3924943HNG3922383HNG3919426HNG3940166HNG3941349HNG3944627HNG3956071HNG3956255HNG3944080HNG3957029HNG4016482HNG4023471HNG4028592HNG4001339HNG4041278HNG4090250HNG4068352HNG4102363HNG4127855HNG4132413HNG4169591HNG4177207HNG4188937HNG4196659HNG4198913HNG4185996HNG4194032HNG2961968HNG3183875HNG3364711HNG3485451HNG3659840HNG3723944HNG3760420HNG3851650HNG3873270HNG3873916HNG3919931HNG3919276HNG3922916HNG3944526HNG3944883HNG3967996HNG3997414HNG4027677HNG4068541HNG4074551HNG4007627HNG4090727HNG4166453HNG4174861HNG4186466HNG4185395HNG3382971HNG3481967HNG3467038HNG3810654HNG3809624HNG3847661HNG3873897HNG3924866HNG4019072HNG4041335HNG3999977HNG4090264HNG4165104HNG4163851HNG4213316HNG4194145HNG4208491HNG4213235HNG3485962HNG3663692HNG3759049HNG3927116HNG3946843HNG4102485HNG4086520HNG4171232HNG4182306HNG4188842HNG4194893HNG4185530HNG3147620HNG3184692HNG3353641HNG3548823HNG3564359HNG3588804HNG3659473HNG3685910HNG3748805HNG3814075HNG3840461HNG3860668HNG3895329HNG3935978HNG3937049HNG3948288HNG3923677HNG4083102HNG4099861HNG4082989HNG4107594HNG4106821HNG4171004HNG4176911HNG4196531HNG3243700HNG3246406HNG3371614HNG3661271HNG3683626HNG3673177HNG3694706H

What the firm is doing

On 11/02/2025, the firm sent an initial "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that an issue was identified which may cause the MiniMed 780G pump to display an alarm indicating insulin delivery is being stopped (Pump error 43 or 41). Customers are instructed to: 1.Respond to all alerts/alarms from your pump. 2.Be prepared to complete the Reservoir & Set procedure after responding to a Pumperror 43 or Pump error 41 alarm. 3.Check that your Alert on High is set during the night or at times when you may not be actively monitoring the pump, making it easier to detect when insulin delivery remains suspended. 4. Keep the pump s audio enabled, making it easier to detect when insulin delivery remains suspended after an alarm is acknowledged. For questions or further assistance, contact Medtronic 24-Hour Technical Support line at 1-800-646-4633, option 1.

DistributionShow details

Worldwide - U.S. National distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The country of Israel.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.