Class IIOngoingZ-1240-2025

GE Healthcare (China) Co., Ltd. recalls Brivo MR355

GE Healthcare (China) Co., Ltd.Beijing, ChinaReported Mar 5, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Brivo MR355, NMRI system

Lot / code information

UDI
) System IDs 082427010096 082427020046 082427040126 082427070076 082427070077 082427070170 082427100245 082427120151 082427180030 082427210147 082427210148 082427220025 082427220045 082427230060 0824…Show all
) System IDs 082427010096 082427020046 082427040126 082427070076 082427070077 082427070170 082427100245 082427120151 082427180030 082427210147 082427210148 082427220025 082427220045 082427230060 082427240060 082427310104 0856270058 0856270059 MRR10200 MRR9920 RU1063MR03 RU8902MR01 (Updated 12/29/25 - Additional System ID 83027805544513) (Updated 1/28/26 - Additional System IDs: 82427220047 82427240055 82427070163 82427060068 82427140137 82427010122 82427280018) (Updated 2/19/26: Additional System ID EG1736MR02)

What the firm is doing

On January 31, 2025, the firm notified customers through Urgent Medical Device Correction letters. Customers were provided with instructions for continued use of their MR systems. All potential users should be made aware of this safety notification and recommended actions. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

DistributionShow details

Domestic US distribution nationwide. International distribution worldwide.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.