Skip to content
RecallWatchMedical Device Safety
Class IOngoingZ-1246-2025

Micro Therapeutics, Inc. recalls Medtronic Pipeline Vantage Embolization Device with Shield Technology

Micro Therapeutics, Inc.Irvine, CA, United StatesReported Mar 12, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Medtronic Pipeline Vantage Embolization Device with Shield Technology27 codes
  • PED3-021-350-16
  • PED3-021-250-14
  • PED3-021-350-20
  • PED3-021-250-20
  • PED3-021-300-16
  • PED3-021-300-20
  • PED3-021-275-12
  • PED3-021-275-16
  • PED3-021-325-12
  • PED3-021-325-14
  • PED3-021-325-16
  • PED3-021-325-20
  • PED3-021-250-12
  • PED3-021-300-14
  • PED3-021-250-16
  • PED3-021-300-12
  • PED3-021-275-14
  • PED3-021-350-14
  • PED3-021-350-25
  • PED3-021-350-12
  • PED3-021-250-10
  • PED3-021-275-20
  • PED3-021-350-18
  • PED3-021-325-18
  • PED3-021-275-18
  • PED3-021-250-18
  • PED3-021-300-18

and Instructions for Use, Part: M056989CDOC2

Lot / code information

Lot #
PED3-021-350-16/00847536030701, 00847536035195, 00847536036529/B317460, B327449, B335184, B340948 — +588 moreShow all
PED3-021-350-16/00847536030701, 00847536035195, 00847536036529/B317460, B327449, B335184, B340948, B349032, B389162, B392918, B424454, B430015, B495821, B497938, B547396, B550307, B551615, B559033, B603377, B623564, B656863, B656887, B658904, B692346, B715057, B715058, B715541, B726894, B735821, B735823, B756029, B756030, B760210, B760211, B768202, B768387, B769453, B770633, B807494; PED3-021-250-14/00847536030411, 00847536034907, 00847536036239/B317936 B332767, B333492, B333536, B334545, B376967, B388041, B391185, B399354, B416713, B431395, B452416, B457963, B467691, B476239, B564173, B564970, B574472, B580923, B629372, B630998, B670305, B673035, B699576, B707568, B709193, B736698, B736893, B739090, B739091, B795110, B796586, B800827, B801770; PED3-021-350-20/00847536030725, 00847536035218, 00847536036543/B319436, B336571, B379993, B383427, B389708, B400103, B400599, B403612, B407062, B415211, B446722, B483987, B490972, B492403, B494049, B552487, B559741, B591782, B591864, B592074, B622656, B626023, B648708, B684905, B715692, B719565, B719567, B719659, B726923, B728926, B735667, B756330, B765053, B768231, B770237, B770248, B792630; PED3-021-250-20/00847536030442, 00847536034938, 00847536036260/B321465, B337378, B360749, B371070, B412868, B467843, B477487, B488713, B548244, B557602, B644034, B673388, B679822, B684208, B686727, B695486, B698571, B703035, B708080, B709204, B711078, B746067, B757056, B757060, B761030, B771498, B775215, B789892, B790055; PED3-021-300-16/00847536030541, 00847536035034, 00847536036369/B321725, B338497, B405720, B418200, B570723, B574473, B603375, B604477, B656277, B658580, B659806, B666425, B681289, B691752, B699606, B703209, B707982, B711365, B725679, B736659, B738114, B769551, B769582, B798864; PED3-021-300-20/00847536030565, 00847536035058, 00847536036383/B322263, B336066, B347327, B376257, B405892, B414535, B458672, B458776, B469040, B490920, B495292, B563197, B607500, B607501, B635214, B636992, B656299, B656355, B657960, B668529, B678482, B680045, B681190, B712110, B712111, B718522, B720923, B725457, B732895, B735956, B737156, B737367, B748272, B748683, B749541, B775230, B778774, B780044; PED3-021-275-12/00847536030466, 00847536034952, 00847536036284/B323044, B383016, B391275, B399486, B454385, B468406, B623580, B623583, B624424, B657776, B666498, B669520, B670179, B670226, B710976, B743820, B743822, B754659, B762034, B768199, B798851; PED3-021-275-16/00847536030480, 00847536034976, 00847536036307/B323046, B361987, B406342, B414102, B458076, B468915, B477868, B558538, B562688, B563304, B603078, B654235, B656399, B657059, B666486, B669610,B670343, B677246, B699308, B736874, B743098, B755194, B755196, B755673, B769461, B771317, B775167, B775168; PED3-021-325-12/00847536030602, 00847536035096, 00847536036420/B324297, B407597, B418843, B494876, B561787, B592073, B623579, B629046, B676507, B678997, B716220, B741395, B751748, B787015, B787017; PED3-021-325-14/00847536030619, 00847536035102, 00847536036437/B324809, B347993, B362685, B459598, B558177, B563209, B575964, B580901, B628946, B675724, B678948, B681904, B714342, B714344, B733352, B733353, B769593, B790478; PED3-021-325-16/00847536030626, 00847536035119, 00847536036444/B324954, B339033, B371691, B418451, B460439, B495941, B603378, B604475, B623578, B624422, B660509, B664402, B666503, B667510, B667920, B668523, B680717, B681747, B733912, B739722, B739723, B739724, B746618, B794191, B794192; PED3-021-325-20/00847536030640, 00847536035133, 00847536036468/B325613, B371743, B418905, B421650, B463895, B487451, B495368, B591952, B591955, B644630, B644769, B644809, B657857, B659135, B659138, B659243, B684852, B716733, B716735, B741305, B741316, B741321, B787131; PED3-021-250-12/00847536030404, 00847536034891, 00847536036222/B326236, B335270, B387931, B390988, B394283, B421249, B477292, B497244, B578963, B664457, B666521, B667945, B669726, B696189, B719158, B724199, B724200, B724764, B741102, B746521, B764797, B764798, B787764, B787765, B787766, B793905; PED3-021-300-14/00847536030534, 00847536035027, 00847536036352/B327828, B335957, B346789, B393640, B399293, B403772, B456576, B547554, B568649, B575299, B631125, B631126, B649258, B650123, B652886, B657773, B683152, B711496, B712157, B712158, B732804, B732807, B754755, B754756, B754757, B760951, B760953, B768142, B769947, B789256; PED3-021-250-16/00847536030428, 00847536034914, 00847536036246/B328095, B416673, B452284, B478579, B494048, B565441, B573470, B575302, B628868, B628869, B630965, B656206, B666535, B667323, B668743, B739139, B739405, B739408, B754155, B754886, B754888, B755651, B794406, B794408, B794426; PED3-021-300-12/00847536030527, 00847536035010, 00847536036345/B338969, B345449, B346121, B361988, B405158, B418358, B454923, B551608, B561782, B562698, B563303, B575894, B677108, B677319, B699289, B703245, B718483, B718491, B739905, B744728, B760716, B766588; PED3-021-275-14/00847536030473, 00847536034969, 00847536036291/B343142, B361366, B379727, B429213, B448933, B452378, B461124, B601345, B607550, B624425, B626121, B678023, B680119, B711192, B713287, B724745, B744169, B744175, B744176, B770971, B770988; PED3-021-350-14/00847536030695, 00847536035188, 00847536036512/B349031, B372357, B388981, B391924, B399969, B421758, B445867, B491603, B492310, B497318, B596593, B629649, B630549, B631109, B670495, B678371, B679957, B713023, B713796, B740630, B740631, B742526, B743935, B753618; PED3-021-350-25/00847536030732, 00847536035225, 00847536036550/B364361, B393438, B403082, B419021, B422233, B430531, B437842, B447224, B488266, B616574, B624414, B629444, B629660, B672265, B672892, B673364, B717131, B717132, B724110, B733587, B733590, B741976, B761681, B768780, B771211, B772034, B773456, B773457; PED3-021-350-12/00847536030688, 00847536035171, 00847536036505/B368529, B406425, B419491, B490325, B496666, B497114, B611908, B615889, B616569, B643508, B643509, B644619, B657109, B658427, B659171, B690687, B715610, B715612, B716621, B742378, B746455

What the firm is doing

On 1/30/2025, Medtronic Neurovascular mailed "Urgent Safety and Labeling Notification" recall notices to healthcare professions who were asked to do the following: Removal (retrieval) is isolated to unused inventory of the Pipeline Vantage 027 devices: 1) Remove and quarantine all unused impacted products in your inventory. 2) Return the impacted products to firm. Firm representative can assist with identifying suitable replacement product. Correction of Instructions for Use applies to Pipeline Vantage 021 devices: The key updates are: 1) Appropriate device diameter and length selection to account for complex anatomy. 2) Techniques to deploy Pipeline Vantage compared to Pipeline Shield using a balance of device tension and compression to achieve adequate wall apposition and landing around curves. 2) Warnings about the consequences of incomplete wall apposition and suboptimal deployment and the increased risk of braid deformation in females, especially in females less-than-or-equal-to 45 years of age. Updated IFU M067713CDOC2 Rev. A (01/2025) was sent with the notice and can be found at: https://manuals.medtronic.com/manuals/ Share this communication within your organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action. Please maintain a copy of this letter for your records and the records of your patients. Complete and return the customer confirmation form via email to neuro.quality@medtronic.com Questions can be directed to the firm at rs.nvquality@medtronic.com or call at 1(800) 633-8766 (US Toll free) or (763) 514-4000 (Worldwide), available 24 hours a day, 7 days a week.

DistributionShow details

Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Republic Of Moldova, Montenegro, Morocco, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2624-2026

Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…

  • Pulmonary Valve Prosthesis Percutaneously Delivered
  • Process control
Medtronic Heart Valves…CAJul 8, 2026
Class IOngoingZ-2622-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026
Class IOngoingZ-2621-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026