Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-1246-2026

Vortex Surgical Inc. recalls Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated…

Vortex Surgical Inc.Saint Charles, MO, United StatesReported Feb 11, 2026 · 8 months ago
Legal News Analyst ·

Reason for recall

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G Laser Probe OS4 Intuitive Exendable VK401323 4. 25G Laser Probe OS4 Flex-Tip VK401115 5. 25G Laser Probe OS4 Illuminated Flex-Tip VK401215 6. 25G Laser Probe OS4 Intuitive Exendable VK401325 7. 27G Laser Probe OS4 Intuitive Exendable VK401327

Lot / code information

UDI
810123483716 Box
UDI
810123483792
Lot #
2410031 2. 23G Laser Probe OS4 Illuminated Flex-Tip Pouch
UDI
810123483730 Box
UDI
810123483815
Lot #
2410034 3. 23G Laser Probe OS4 Intuitive Exendable Pouch
UDI
810123483761 Box
UDI
810123483846
Lot #
2410033 4. 25G Laser Probe OS4 Flex-Tip Pouch
UDI
810123483723 Box
Show 11 more code fields
UDI
810123483808
Lot #
2410032 2410058 2411022; 5. 25G Laser Probe OS4 Illuminated Flex-Tip Pouch
UDI
810123483747 Box
UDI
810123483822
Lot #
2410035 6. 25G Laser Probe OS4 Intuitive Exendable Pouch
UDI
810123483778 Box
UDI
810123483853
Lot #
2412025 7. 27G Laser Probe OS4 Intuitive Exendable Pouch
UDI
810123483785 Box
UDI
810123483860
Lot #
2411021

What the firm is doing

On December 16, 2025, the firm began notifying customers via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were instructed to discontinue use of affected product and to return product for Vortex Surgical to evaluate the seals; use UPS account 23623Y to ship unused products to Vortex Surgical Inc for replacement or refund at: Attn: Andrew Ritts Recall Tyvek Seal, Vortex Surgical Inc., 4 Research Park Dr., Suite 124, St. Charles, MO 63304; complete and return the Acknowledgement and Receipt Form via mail or email to: ARitts@VortexSurgical.com. Actions have been implemented to improve training of individuals associated with sealing and inspecting seals to prevent reoccurrence. The sealer associated with the issue is being updated to include a guard to prevent reoccurrence. If you have any questions, contact the Director of Regulatory Affairs at 635-778-4350 or email to: ARitts@VortexSurgical.com from 8am to 5pm.

DistributionShow details

Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.