Kung Shin Plastics Co. Ltd. recalls CircuitGuard ThermoFlo Filter -Combined Heat and Moisture Exchanger

Kung Shin Plastics Co. Ltd.Taipei Hsien, TaiwanReported Mar 12, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • CircuitGuard ThermoFlo Filter -Combined Heat and Moisture Exchanger, Bacterial Viral Filter w O2 Port and Elbow (Model/Catalogue number: SK200CPE / 6126)
    Affected lot
    C2O24002

What the firm is doing

On January 9, 2025, the firm sent a recall notice to their sole distributor in the US, followed by a more detailed notice on January 10. Customers were instructed to locate and immediately discontinue use of affected products. Affected product should be quarantined and returned. If products of affected batches have been sold, the distributor should immediately notify the buyers to stop using them, quarantine them, and destroy them.

DistributionShow details

US Nationwide distribution in the state of IL.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.