Class IIOngoingZ-1297-2025

GE Medical Systems, LLC recalls GE Healthcare Revolution Apex

GE Medical Systems, LLCWaukesha, WI, United StatesReported Mar 19, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed

Lot / code information

UDI
(01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/
Serial #
REV2X2300082CN; System ID Number: 650493REV1
UDI
To be provided, Mfg. Lot/
Serial #
REVVX1900035CN; System ID Number: 630266CTRV
UDI
(01)00195278460271(11)240800(21)REV2X2400093CN, Mfg. Lot/
Serial #
REV2X2400093CN; System ID Number: 773702REVCT6ED
UDI
(01)00195278460271(11)231000(21)REV2X2300071CN, Mfg. Lot/
Serial #
REV2X2300071CN; System ID Number: 312996CTRV2
UDI
(01)00195278460271(11)241000(21)REV2X2400126CN, Mfg. Lot/
Serial #
REV2X2400126CN; System ID Number: 1000000018
Show 102 more code fields
UDI
To be provided, Mfg. Lot/
Serial #
REV2X2300074CN; System ID Number: 617726Y6REV2
UDI
(01)00840682118552(11)170900(21)REVVX1700090CN, Mfg. Lot/
Serial #
REVVX1700090CN; System ID Number: 617726Y6REV1
UDI
(01)00840682118552(11)160700(21)REVVX1600018CN, Mfg. Lot/
Serial #
REVVX1600018CN; System ID Number: 336890APEX
UDI
(01)00195278460271(11)240900(21)REV2X2400116CN, Mfg. Lot/
Serial #
REV2X2400116CN; System ID Number: 704801APEX
UDI
(01)00195278460271(11)241000(21)REV2X2400139CN, Mfg. Lot/
Serial #
REV2X2400139CN; System ID Number: 336832APEX
UDI
(01)00195278460271(11)241000(21)REV2X2400129CN, Mfg. Lot/
Serial #
REV2X2400129CN; System ID Number: 980993CT1
UDI
(01)00840682146616(11)210700(21)REV2A2100029CN, Mfg. Lot/
Serial #
REV2A2100029CN; System ID Number: 910715REV1
UDI
(01)00195278460271(11)241200(21)REV2X2400155CN, Mfg. Lot/
Serial #
REV2X2400155CN; System ID Number: 603650APEX1
UDI
(01)00195278460271(11)240600(21)REV2X2400068CN, Mfg. Lot/
Serial #
REV2X2400068CN; System ID Number: 603526APEX
UDI
(01)00195278469700(11)230100(21)REV242200004CN, Mfg. Lot/
Serial #
REV242200004CN; System ID Number: 973971REV161
UDI
(01)00195278460271(11)241200(21)REV2X2400158CN, Mfg. Lot/
Serial #
REV2X2400158CN; System ID Number: 914787APEX
UDI
(01)00195278460271(11)231000(21)REV2X2300081CN, Mfg. Lot/
Serial #
REV2X2300081CN; System ID Number: 718963EDCT
UDI
(01)00195278460271(11)240600(21)REV2X2400070CN, Mfg. Lot/
Serial #
REV2X2400070CN; System ID Number: 315473APEX1
UDI
(01)00195278460271(11)241100(21)REV2X2400152CN, Mfg. Lot/
Serial #
REV2X2400152CN; System ID Number: 614RMHAPEX1
UDI
(01)00195278460271(11)240300(21)REV2X2400026CN, Mfg. Lot/
Serial #
REV2X2400026CN; System ID Number: 724357APEX
UDI
(01)00195278460271(11)230800(21)REV2X2300046CN, Mfg. Lot/
Serial #
REV2X2300046CN; System ID Number: 843797REVCT2
UDI
(01)00840682123471(11)211100(21)REVV82100041CN, Mfg. Lot/
Serial #
REVV82100041CN; System ID Number: 806775APEX
UDI
(01)00195278460271(11)231000(21)REV2X2300075CN, Mfg. Lot/
Serial #
REV2X2300075CN; System ID Number: 703776APEX
UDI
To be provided, Mfg. Lot/
Serial #
REV2X2400167CN; System ID Number: 414805REVCT
UDI
t Applicable, Mfg. Lot/
Serial #
00000441066CN8; System ID Number: 262577REVCT
UDI
(01)00195278460271(11)240500(21)REV2X2400055CN, Mfg. Lot/
Serial #
REV2X2400055CN; System ID Number: 832CT000059
UDI
(01)00840682146616(11)220100(21)REV2A2200006CN, Mfg. Lot/
Serial #
REV2A2200006CN; System ID Number: 833CT000155
UDI
(01)00195278460271(11)240500(21)REV2X2400053CN, Mfg. Lot/
Serial #
REV2X2400053CN; System ID Number: AH7084CT02
UDI
(01)00840682146616(11)220100(21)REV2A2200007CN, Mfg. Lot/
Serial #
REV2A2200007CN; System ID Number: AH1923CT01
UDI
(01)00195278460271(11)240300(21)REV2X2400024CN, Mfg. Lot/
Serial #
REV2X2400024CN; System ID Number: BG4549CT04
UDI
(01)00195278460271(11)230600(21)REV2X2300026CN, Mfg. Lot/
Serial #
REV2X2300026CN; System ID Number: E416249APEXCTE
UDI
(01)00195278460271(11)231100(21)REV2X2300087CN, Mfg. Lot/
Serial #
REV2X2300087CN; System ID Number: 082421140968
UDI
To be provided, Mfg. Lot/
Serial #
t Available; System ID Number: 082421140960
UDI
(01)00195278633149(11)240500(21)CBFCG2400006HM, Mfg. Lot/
Serial #
CBFCG2400006HM; System ID Number: 082421140959
UDI
(01)00195278633149(11)240500(21)CBFCG2400004HM, Mfg. Lot/
Serial #
CBFCG2400004HM; System ID Number: DK1044CT12
UDI
(01)00840682146616(11)200200(21)REV2A2000001CN, Mfg. Lot/
Serial #
REV2A2000001CN; System ID Number: DK1547CT01
UDI
(01)00840682123471(11)210100(21)REVV82100002CN, Mfg. Lot/
Serial #
REVV82100002CN; System ID Number: DK1051CT04
UDI
To be provided, Mfg. Lot/
Serial #
00000442412CN3; System ID Number: CT52261701
UDI
To be provided, Mfg. Lot/
Serial #
REV2X2300050CN; System ID Number: M40230104
UDI
(01)00840682118552(11)160600(21)REVVX1600012CN, Mfg. Lot/
Serial #
REVVX1600012CN; System ID Number: REV020
UDI
To be provided, Mfg. Lot/
Serial #
00000442334CN9; System ID Number: CT00486
UDI
(01)00195278460271(11)231000(21)REV2X2300079CN, Mfg. Lot/
Serial #
REV2X2300079CN; System ID Number: KRCT54972221
UDI
To be provided, Mfg. Lot/
Serial #
t Available; System ID Number: KRCT54893311
UDI
To be provided, Mfg. Lot/
Serial #
t Available; System ID Number: 0850211084
UDI
To be provided, Mfg. Lot/
Serial #
REV2X2300016CN; System ID Number: NO1068CT04
UDI
(01)00840682118552(11)171000(21)REVVX1700106CN, Mfg. Lot/
Serial #
REVVX1700106CN; System ID Number: P00652CT01
UDI
(01)00195278460271(11)240200(21)REV2X2400017CN, Mfg. Lot/
Serial #
REV2X2400017CN; System ID Number: PL1672CT03
UDI
(01)00840682146616(11)210600(21)REV2A2100019CN, Mfg. Lot/
Serial #
REV2A2100019CN; System ID Number: SA1420CT01
UDI
To be provided, Mfg. Lot/
Serial #
REV2X2400137CN; System ID Number: SA2973CT01
UDI
To be provided, Mfg. Lot/
Serial #
REV2X23000030CN; System ID Number: CT53972381
UDI
To be provided, Mfg. Lot/
Serial #
REV2X2400121CN; System ID Number: A201CT07
UDI
(01)00195278460271(11)240200(21)REV2X2400014CN, Mfg. Lot/
Serial #
REV2X2400014CN; System ID Number: AC001CT05
UDI
(01)00840682118552(11)181000(21)REVVX1800101CN, Mfg. Lot/
Serial #
REVVX1800101CN; System ID Number: 541010CT02
UDI
(01)00840682118552(11)190900(21)REVVX1900106CN, Mfg. Lot/
Serial #
REVVX1900106CN; System ID Number: THCTX24524601
UDI
(01)00195278460271(11)241000(21)REV2X2400131CN, Mfg. Lot/
Serial #
REV2X2400131CN; System ID Number: AE1025CT03
UDI
(01)00840682118552(11)210300(21)REVVX2100010CN, Mfg. Lot/
Serial #
REVVX2100010CN

What the firm is doing

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/18/2025 via letter delivered using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your system. Please monitor the floor near the base of the gantry for any accumulation of blue fluid. If blue fluid is observed, wear nitrile or rubber gloves, clean up the fluid by absorbing it with a noncombustible absorbent material, and dispose of the absorbent material in a suitable container for disposal. Please contact your local GE HealthCare service representative if any blue fluid is observed. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please complete and return the attached acknowledgement form to recall.25504@gehealthcare.com."

DistributionShow details

Worldwide distribution.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.