Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha recalls Brand Name: Hitachi Proton Beam Therapy System Ref: PROBEAT-FR
Reason for recall
Software anomaly in the patient positioning system may result in positional discrepancy.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Brand Name: Hitachi Proton Beam Therapy System Ref:PROBEAT-FRAffected lotPTFR-01/
What the firm is doing
Between the dates of February 5-11, 2026, Hitachi, Ltd, issued an "URGENT: MEDICAL DEVICE RECALL" notification dated January 21, 2026, via Service Team Manager to affected consignees. Hitachi is asking consignees to take the following actions: 1. Until the PPS software has been updated with the corrected version, users are instructed to follow the System Operation instruction in the document. 2. Users are also asked to post the document in a prominent space and sign its cover. 3. Complete and return the acknowledgment and receipt form. Action by the firm: Hitachi plans to update the PPS software to a corrected version that prevents the mismatch between the commanded error correction parameter and the internal variable. starting in late March 2026 and the users will be contacted after the update. If you have any questions, please call QA Manager at Tel: (408) 986-6300. Monday through Friday, 8:00 AM to 5:00 PM, Central Time.
DistributionShow detailsHide
US distribution to states of: TX, Washington DC
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1429-2026
- FDA 510(k) clearance · K232032The device's official FDA premarket clearance record
- FDA device classification · LHNOfficial FDA classification for this device type
- CFR regulation · 21 CFR 892.5050The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Hitachi, Ltd. Radiation Oncology Systems, KashiwanohaSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) BASIC PACK
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Eye Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) KIT
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Orthopedic Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) CHI STD NEURO CRANI OPT 2
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Neurological Tray
- Nonconforming Material/Component