Stryker Corporation recalls Nasopore FD fragmentable nasal dressing
Reason for recall
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Nasopore FD fragmentable nasal dressing, Catalog Numbers 5400-020-108 and 5400-020-108ITL
Lot / code information
- UDI
- 07613327117226
- Lot #
- 2023011317 2023021717 2023030932 2023040317 2023083017 2023102517 2024020517 2024080717; 5400-020-108ITL
- UDI
- 07613327117226
- Lot #
2023012017 2023032817 2023042017 2023091117 2023111417 2023120517 2024011817 2023020732 2023041117 2023062017 2023092017 2023112017 2024010817 2024012417 2023031717 2023041417 2023082317 2023103117 20…Show all
2023012017 2023032817 2023042017 2023091117 2023111417 2023120517 2024011817 2023020732 2023041117 2023062017 2023092017 2023112017 2024010817 2024012417 2023031717 2023041417 2023082317 2023103117 2023112817 2024011217
What the firm is doing
An Urgent Medical Device Recall notification letter dated 2/24/25 was sent to customers. Actions to be taken 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. Instruments.recalls@stryker.com. RESPONSE IS REQUIRED.
DistributionShow detailsHide
US Nationwide distribution.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1476-2025
- FDA 510(k) clearance · K070715The device's official FDA premarket clearance record
- FDA device classification · NHBOfficial FDA classification for this device type
- CFR regulation · 21 CFR 874.3620The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Stryker CorporationSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm