Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-1552-2025

CareFusion 303, Inc. recalls MEDFLEX

CareFusion 303, Inc.San Diego, CA, United StatesReported Apr 23, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

MEDFLEX1 code
  • 1119-00
MEDBANK MAIN1 code
  • 1137-00
MEDFLEX 2.01 code
  • 1139-00
MEDBANK TOWER MAIN1 code
  • 1145-00
MEDBANK MINI MAIN2 codes
  • 1147-00 BD PYXIS MEDBANK Facility Software
  • 139090-01
BD PYXS MEDBNK TWR MN 2HH1FH2HM2FM2FG1FF1 code
  • 169-100
BD PYXIS MEDBANK MINI 1HH-1FH- 1HM1 code
  • 169-101
BD PYXIS MEDBANK MINI 1HH-1HM- 1FM1 code
  • 169-102
BD PYXIS MEDBANK MINI 1HH-3HM1 code
  • 169-103
BD PYXIS MEDBANK MINI 2FH1 code
  • 169-104
BD PYXIS MEDBANK MINI 2FM1 code
  • 169-105
BD PYXIS MEDBANK MINI 2HH-1FH1 code
  • 169-106
BD PYXIS MEDBANK MINI 2HH-1FM1 code
  • 169-107
BD PYXIS MEDBANK MINI 2HH-2HM1 code
  • 169-108
BD PYXIS MEDBANK MINI 2HM-1FM1 code
  • 169-109
BD PYXIS MEDBANK MINI 3HH-1FG1 code
  • 169-110
BD PYXIS MEDBANK MINI 3HH-1HM1 code
  • 169-111
BD PYXIS MEDBANK MINI 4HH1 code
  • 169-112
BD PYXIS MEDBANK MINI 4HM1 code
  • 169-113
BD PYXIS MDBANK TWR MN 7HH-1HM-3FM RxNow1 code
  • 169-132
BD PYXIS MEDBANK MINI 1FH-1FM1 code
  • 169-137
BD PYXIS MEDBANK TWR MN 1FH- 10HM-1FM1 code
  • 169-138
BD PYXIS MEDBANK TWR MN 2HH- 4FH-2FM1 code
  • 169-139
OBS BD PYXIS MB TOWER MAIN1 code
  • 169-153
OBS BD PYXIS MB TWR RXNOW MAIN W/MED PRI1 code
  • 169-155
OBS BD PYXIS MB TWR RXNOW MAIN W/PAT PRI1 code
  • 169-156
OBS BD PYXIS MB MINI MAIN W/CUBIE DRWS1 code
  • 169-159
OBS BD PYXIS MB MINI MAIN W/MATRIX DRWS1 code
  • 169-160
OBS BD PYXIS MB MINI MAIN W/HH FALSE DRW1 code
  • 169-161
OBS BD PYXIS MB MINI MAIN W/FH FALSE DRW1 code
  • 169-162
BD PYXIS MEDBANK TWR MN 10HH- 1FH-1FM1 code
  • 169-23
BD PYXIS MEDBANK TWR MN 10HH- 2FH1 code
  • 169-24
BD PYXIS MEDBANK TWR MN 10HH- 2FM1 code
  • 169-25
BD PYXIS MEDBANK TWR MN 12HH- 1FF1 code
  • 169-26
BD PYXIS MEDBANK TWR MN 12HH- 1FH1 code
  • 169-27
BD PYXIS MEDBANK TWR MN 12HH- 1FM1 code
  • 169-28
BD PYXIS MEDBANK TWR MN 14HH1 code
  • 169-29
BD PYXIS MEDBANK TWR MN 1HH- 1HM-2FM-4FF1 code
  • 169-30
BD PYXIS MEDBANK TWR MN 2FH- 2HM-4FM1 code
  • 169-31
BD PYXIS MEDBANK TWR MN 2HH- 10HM-1FM1 code
  • 169-32
BD PYXIS MEDBANK TWR MN 2HH- 1FH-5FM1 code
  • 169-33
BD PYXIS MEDBANK TWR MN 2HH- 2FH-4FM1 code
  • 169-34
BD PYXIS MEDBANK TWR MN 2HH- 2FM-4FF1 code
  • 169-35
BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FF1 code
  • 169-36
BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FM1 code
  • 169-37
BD PYXIS MEDBANK TWR MN 2HH- 6FM1 code
  • 169-38
BD PYXIS MEDBANK TWR MN 2HH- 6HM-3FM1 code
  • 169-39
BD PYXIS MEDBANK TWR MN 2HH- 8HM-2FM1 code
  • 169-40
BD PYXIS MEDBANK TWR MN 2HM- 6FM1 code
  • 169-41
BD PYXIS MEDBANK TWR MN 3FH-4FM1 code
  • 169-42
BD PYXIS MEDBANK TWR MN 3HH- 7HM-2FM1 code
  • 169-43
BD PYXIS MEDBANK TWR MN 4HH- 10HM1 code
  • 169-44
BD PYXIS MEDBANK TWR MN 4HH- 1FH-2HM-3FM1 code
  • 169-45
BD PYXIS MEDBANK TWR MN 4HH- 1FH-4FM1 code
  • 169-46
BD PYXIS MEDBANK TWR MN 4HH- 1FH-4HM-2FM1 code
  • 169-47
BD PYXIS MEDBANK TWR MN 4HH- 1FH-6HM-1FM1 code
  • 169-48
BD PYXIS MEDBANK TWR MN 4HH- 2FH-2HM-2FM1 code
  • 169-49
BD PYXIS MEDBANK TWR MN 4HH- 2FH-3FM1 code
  • 169-50
BD PYXIS MEDBANK TWR MN 4HH- 2HM-4FM1 code
  • 169-51
BD PYXIS MEDBANK TWR MN 4HH- 3FH-2FM1 code
  • 169-52
BD PYXIS MEDBANK TWR MN 4HH- 4FH-1FM1 code
  • 169-53
BD PYXIS MEDBANK TWR MN 4HH- 4HM-3FM1 code
  • 169-54
BD PYXIS MEDBANK TWR MN 4HH-5FH1 code
  • 169-55
BD PYXIS MEDBANK TWR MN 4HH- 5FM1 code
  • 169-56
BD PYXIS MEDBANK TWR MN 4HH- 6HM-2FM1 code
  • 169-57
BD PYXIS MEDBANK TWR MN 4HH- 8HM-1FM1 code
  • 169-58
BD PYXIS MEDBANK TWR MN 4HM- 5FM1 code
  • 169-59
BD PYXIS MEDBANK TWR MN 5HH- 3HM-3FM1 code
  • 169-60
BD PYXIS MEDBANK TWR MN 5HH- 9HM1 code
  • 169-61
BD PYXIS MEDBANK TWR MN 6FH-1FM1 code
  • 169-62
BD PYXIS MEDBANK TWR MN 6HH- 1FH-2HM-2FM1 code
  • 169-63
BD PYXIS MEDBANK TWR MN 6HH- 1FH-3FM1 code
  • 169-64
BD PYXIS MEDBANK TWR MN 6HH- 2FH-2FM1 code
  • 169-65
BD PYXIS MEDBANK TWR MN 6HH- 2FH-2HM-1FM1 code
  • 169-66
BD PYXIS MEDBANK TWR MN 6HH-4FH1 code
  • 169-67
BD PYXIS MEDBANK TWR MN 6HH- 4FM1 code
  • 169-68
BD PYXIS MEDBANK TWR MN 6HH- 4HM-2FM1 code
  • 169-69
BD PYXIS MEDBANK TWR MN 6HH- 6HM-1FM1 code
  • 169-70
BD PYXIS MEDBANK TWR MN 6HH- 8HM1 code
  • 169-71
BD PYXIS MEDBANK TWR MN 7FH1 code
  • 169-72

BD PYXIS M

Lot / code information

UDI
1119-00/10885403512704; 1137-00; 1139-00/10885403512704; 1145-00/10885403517167; 1147-00/10885403517181; 139090-01; 169-100/10885403512568; 169-101/10885403512544; 169-102/10885403512544; 169-103/1088…Show all
1119-00/10885403512704; 1137-00; 1139-00/10885403512704; 1145-00/10885403517167; 1147-00/10885403517181; 139090-01; 169-100/10885403512568; 169-101/10885403512544; 169-102/10885403512544; 169-103/10885403512544; 169-104/10885403512544; 169-105/10885403512544; 169-106/10885403512544; 169-107/10885403512544; 169-108/10885403512544; 169-109/10885403512544; 169-110/10885403512544; 169-111/10885403512544; 169-112/10885403512544; 169-113/10885403512544; 169-132/10885403512568; 169-137/10885403512544; 169-138/10885403512568; 169-139/10885403512568; 169-153/10885403512568; 169-155/10885403512568; 169-156/10885403512568; 169-159/10885403512544; 169-160/10885403512544; 169-161/10885403512544; 169-162/10885403512544; 169-23/10885403512568; 169-24/10885403512568; 169-25/10885403512568; 169-26/10885403512568; 169-27/10885403512568; 169-28/10885403512568; 169-29/10885403512568; 169-30/10885403512568; 169-31/10885403512568; 169-32/10885403512568; 169-33/10885403512568; 169-34/10885403512568; 169-35/10885403512568; 169-36/10885403512568; 169-37/10885403512568; 169-38/10885403512568; 169-39/10885403512568; 169-40/10885403512568; 169-41/10885403512568; 169-42/10885403512568; 169-43/10885403512568; 169-44/10885403512568; 169-45/10885403512568; 169-46/10885403512568; 169-47/10885403512568; 169-48/10885403512568; 169-49/10885403512568; 169-50/10885403512568; 169-51/10885403512568; 169-52/10885403512568; 169-53/10885403512568; 169-54/10885403512568; 169-55/10885403512568; 169-56/10885403512568; 169-57/10885403512568; 169-58/10885403512568; 169-59/10885403512568; 169-60/10885403512568; 169-61/10885403512568; 169-62/10885403512568; 169-63/10885403512568; 169-64/10885403512568; 169-65/10885403512568; 169-66/10885403512568; 169-67/10885403512568; 169-68/10885403512568; 169-69/10885403512568; 169-70/10885403512568; 169-71/10885403512568; 169-72/10885403512568; 169-73/10885403512568; 169-74/10885403512568; 169-75/10885403512568; 169-76/10885403512568; 169-77/10885403512568; 169-78/10885403512568; 169-79/10885403512568; 169-80/10885403512568; 169-81/10885403512568; 139089-01

What the firm is doing

05OCT2023, Customer Release Notes were sent to customers informing them of system release v1.3.4. On 2/18/2025, correction notices were mailed and emailed to Customers, Directors of Pharmacy, and Facility Medication Management Administrators who were informed the following: 1) Customers with system administrator privileges can uncheck the option for the "Clear All" button in the BD Pyxis MedBank Facility Software, which will disable the feature and prevent the button from displaying on the patient profile selection screen. 2) Best practice is to verify the item being removed is accurate against the provider s order, and if available utilize barcode medication administration as an additional check to ensure the correct medication is retrieved prior to administration.1 3) Share and disseminate this notice within your facility network to ensure they are aware of the field action. 4) Complete and return the Customer Response Form via email to BDRC23@bd.com BD Pyxis MedBank Facility Software version 2.5.1.4 within MedBank system release v1.7 is currently available and will address this issue. BD will contact impacted customers to schedule a time for software upgrade. If you have questions contact the firm's Technical Support at 1-866-930-9251 Phone Hours: 24 hours, 7 days a week.

DistributionShow details

US Nationwide distribution in the states of KS, TX, CA, PA, NJ, FL, IN, AZ, IL, MA, NY, KY, UT, MD, WA, OH, AL.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.