Beckman Coulter Mishima K.K. recalls CHEMISTRY ANALYZER AU5800

Beckman Coulter Mishima K.K.Sunto-Gun, JapanReported Mar 18, 2026 · 4 months ago
Legal News Analyst ·

Reason for recall

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

CHEMISTRY ANALYZER AU58005 codes
  • B96697
  • B96698
  • B23279
  • B23280
  • B23281

Lot / code information

REF/UDI-DI/ Newer Instrument Installed with PC preinstalled with Microsoft SQL Server 2022 Express: B96697/14987666543016/2025061442, 2025061446, 2025061452, 2025061454, 2025061464, 2025071479, 2025071488, 2025081513, 2025081516, 2025091536, 2025091538, 2025091543, 2025091559, 2025091561, 2025111598, 2025111602, 2025111604, 2025111605, 2025111606, 2025121618, 2025121622, 2025121623. B96698/14987666543023/2025061449, 2025061451, 2025061457, 2025061467, 2025071475, 2025071476, 2025071478, 2025071481, 2025071495, 2025071497, 2025081503, 2025081524, 2025091539, 2025091549, 2025091551. Older Instrument Replaced with PC preinstalled with Microsoft SQL Server 2022 Express: B23279/14987666541296/2014011316, 2014071647, 2014121858, 2014121906. B23280/14987666541302/2016022593. B23281/14987666541319/5094354. B96697/14987666543016/2019074683. B96698/14987666543023/2017063377, 2019114891, 2024101288

DistributionShow details

US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.