Beckman Coulter Mishima K.K. recalls CHEMISTRY ANALYZER AU5800
Reason for recall
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
CHEMISTRY ANALYZER AU58005 codes
- B96697
- B96698
- B23279
- B23280
- B23281
Lot / code information
REF/UDI-DI/ Newer Instrument Installed with PC preinstalled with Microsoft SQL Server 2022 Express: B96697/14987666543016/2025061442, 2025061446, 2025061452, 2025061454, 2025061464, 2025071479, 2025071488, 2025081513, 2025081516, 2025091536, 2025091538, 2025091543, 2025091559, 2025091561, 2025111598, 2025111602, 2025111604, 2025111605, 2025111606, 2025121618, 2025121622, 2025121623. B96698/14987666543023/2025061449, 2025061451, 2025061457, 2025061467, 2025071475, 2025071476, 2025071478, 2025071481, 2025071495, 2025071497, 2025081503, 2025081524, 2025091539, 2025091549, 2025091551. Older Instrument Replaced with PC preinstalled with Microsoft SQL Server 2022 Express: B23279/14987666541296/2014011316, 2014071647, 2014121858, 2014121906. B23280/14987666541302/2016022593. B23281/14987666541319/5094354. B96697/14987666543016/2019074683. B96698/14987666543023/2017063377, 2019114891, 2024101288
DistributionShow detailsHide
US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1553-2026
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Beckman Coulter Mishima K.K.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) BASIC PACK
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Eye Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) KIT
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Orthopedic Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) CHI STD NEURO CRANI OPT 2
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Neurological Tray
- Nonconforming Material/Component