Class IIOngoingZ-1578-2025

Micro-X Ltd. recalls Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 P…

Micro-X Ltd.Tonsley, AustraliaReported Apr 23, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component:N/A
    UDI MSN0334
    Affected lot
    MSN0334

What the firm is doing

On March 28, 2025, Micro-X Ltd. issued a "Urgent: Medical Device Correction" notification to affected consignees via E-Mail. Micro-X asked consignees to take the following actions: 1. Continue to use your product as normal in accordance with its intended use. 2. Your local service representative will be in contact to apply this update to the systems at the next site visit. 3. Circulate this notice to all users of this device so they are aware of the issue. 4. If the product has been further distributed to other facilities, please ensure this notice is immediately shared with them. 5. Please retain this letter with your system(s) documentation. 6. Please complete and return the attached response form to Micro-X to the Micro-X LTD in Australia: support@micro-x.com and copy djerkovic@micro-x.com.

DistributionShow details

US Nationwide distribution in the states of CA, NJ, and Puerto Rico (US territory).

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.