Class IOngoingZ-1635-2025

GE Medical Systems China Co., Ltd. recalls GE Healthcare CARESTATION 750 A1

GE Medical Systems China Co., Ltd.Wuxi, ChinaReported May 7, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • GE Healthcare CARESTATION 750 A1, Model/REF Number 1012-9750-000. Anesthesia system
    GTIN 00840682145596
    400 affected lots
    SQS24500010MASQS24500011MASQS24500012MASQS24500013MASQS24500014MASQS24500015MASQS24500016MASQS24500017MA
    +392 moreSQS24130011MASQS24130012MASQS24130013MASQS24130014MASQS24130015MASQS24130016MASQS24130017MASQS24130018MASQS24170062MASQS24170063MASQS24170064MASQS24170065MASQS24170066MASQS24170079MASQS24170080MASQS24170084MASQS24170085MASQS24170086MASQS24170087MASQS24170088MASQS24190115MASQS24190116MASQS24190117MASQS24430007MASQS24430008MASQS24430009MASQS24430010MASQS24150018MASQS24390013MASQS24390014MASQS24460010MASQS23210004MASQS23210001MASQS23210002MASQS23210003MASQS25020021MASQS25020022MASQS25020023MASQS25020024MASQS25020025MASQS24480001MASQS24250003MASQS24160054MASQS24160055MASQS24160056MASQS24160057MASQS24180105MASQS24190120MASQS24190121MASQS24190122MASQS24190123MASQS24190124MASQS25020015MASQS24440002MASQS24440003MASQS24440004MASQS24440005MASQS24440006MASQS24440007MASQS24440008MASQS24440009MASQS24440010MASQS24440011MASQS24440012MASQS24440013MASQS24460001MASQS24500002MASQS24150022MASQS24150023MASQS24150024MASQS24150025MASQS24150026MASQS24150027MASQS24150028MASQS24150029MASQS24150030MASQS24150031MASQS24150032MASQS25030026MASQS23340006MASQS23340005MASQS23380004MASQS23380005MASQS24190118MASQS24190119MASQS24490033MASQS24490034MASQS24210005MASQS24210006MASQS24380019MASQS24370002MASQS24370003MASQS24370004MASQS24370005MASQS25030028MASQS25030029MASQS25030030MASQS24420022MASQS24420023MASQS24290048MASQS24290049MASQS24290050MASQS24290051MASQS24290052MASQS24290053MASQS24290054MASQS24290055MASQS24290056MASQS24290057MASQS24290058MASQS24290059MASQS24290060MASQS24290062MASQS24290063MASQS24290064MASQS24290065MASQS24290066MASQS24290067MASQS24290068MASQS24290069MASQS24290070MASQS24290071MASQS24290072MASQS24290073MASQS24290074MASQS24290075MASQS24290076MASQS24290077MASQS24290078MASQS24290079MASQS24480002MASQS24480003MASQS24480004MASQS24480005MASQS24480006MASQS23380003MASQS24320007MASQS24320009MASQS24510016MASQS24390007MASQS24390008MASQS24390009MASQS23360001MASQS23360002MASQS24040003MASQS24040004MASQS24060003MASQS24060004MASQS24060005MASQS24060006MASQS24060007MASQS24060008MASQS24060009MASQS24170060MASQS24170061MASQS24070007MASQS24180095MASQS24150016MASQS24380021MASQS24380022MASQS24380023MASQS24040012MASQS24070008MASQS24070009MASQS24070010MASQS24070011MASQS24120010MASQS24050080MASQS24050081MASQS24050082MASQS24050083MASQS24050084MASQS24050085MASQS24050086MASQS24050087MASQS24420001MASQS24420002MASQS24420003MASQS24420004MASQS24420005MASQS24420006MASQS24420007MASQS24420008MASQS24420009MASQS24420010MASQS24420011MASQS24420012MASQS24420013MASQS24420014MASQS24420015MASQS24420016MASQS24420017MASQS24420018MASQS24420019MASQS24420020MASQS24420021MASQS23430010MASQS24420024MASQS23380006MASQS23380007MASQS23490002MASQS23490003MASQS23490004MASQS23490005MASQS24060024MASQS23440024MASQS24460008MASQS24460009MASQS24460011MASQS24380002MASQS24380003MASQS24380004MASQS24390005MASQS24390006MASQS24210007MASQS24210008MASQS24070002MASQS24070003MASQS24090006MASQS24090007MASQS24040005MASQS24040006MASQS24040007MASQS24040008MASQS24040009MASQS24040010MASQS24400007MASQS24400008MASQS24400009MASQS24400010MASQS24330001MASQS24330002MASQS24330003MASQS24330004MASQS24330005MASQS24330006MASQS25030027MASQS23440023MASQS24010021MASQS24010022MASQS24010023MASQS24010024MASQS24010025MASQS24010026MASQS24010027MASQS24150033MASQS24150034MASQS24150035MASQS24150036MASQS24150037MASQS24150038MASQS24150039MASQS24150040MASQS24320006MASQS23500001MASQS23500004MASQS24320008MASQS24370006MASQS24380020MASQS24170059MASQS23010001MASQS23080033MASQS24110014MASQS24400001MASQS24110013MASQS24320010MASQS24320011MASQS24070004MASQS24070005MASQS24070006MASQS24360001MASQS24360002MASQS24360003MASQS24200002MASQS24400003MASQS23320001MASQS23320002MASQS23320003MASQS23320004MASQS23320005MASQS23320006MASQS23320007MASQS23320008MASQS24170071MASQS24170072MASQS24170073MASQS24170074MASQS24330007MASQS24330008MASQS24330009MASQS24330010MASQS24360006MASQS24360007MASQS24360010MASQS24410001MASQS24490008MASQS24490010MASQS24490015MASQS24490018MASQS24490021MASQS24490022MASQS24490023MASQS24490024MASQS24490025MASQS24490028MASQS24490029MASQS24490030MASQS24490031MASQS24490005MASQS24490006MASQS24490011MASQS24490012MASQS24490014MASQS24490016MASQS24490017MASQS24490019MASQS24490020MASQS24490026MASQS24490027MASQS24340021MASQS24340022MASQS24340023MASQS24310003MASQS24490001MASQS24490002MASQS24490003MASQS24490004MASQS24490007MASQS24490009MASQS24490013MASQS24360004MASQS24360005MASQS24360008MASQS24360009MASQS23430011MASQS24070001MASQS24160042MASQS24160043MASQS24340014MASQS24340015MASQS24340016MASQS24340017MASQS24340018MASQS24340019MASQS24340020MASQS24490032MASQS24140005MASQS24140006MASQS24050036MASQS24050037MASQS24050038MASQS24050039MASQS24050040MASQS24050041MASQS24050042MASQS24050043MASQS24050044MASQS24050045MASQS24040014MASQS24500001MASQS24410002MASQS24410003MASQS24410004MASQS24410005MASQS24410006MASQS24410007MASQS24410008MASQS24410009MASQS24410010MASQS24410011MASQS24410012MASQS24090005MASQS24110005MASQS23430014MASQS23430015MASQS23430016MASQS23430017MASQS23430018MASQS23430019MASQS24180099MASQS24180100MASQS24180101MASQS24180102MASQS24180103MASQS24180104MASQS24470004MASQS24470005MASQS23480009MASQS23480010MASQS24510020MASQS24510029MASQS24510030MA

What the firm is doing

GE Healthcare issued an URGENT MEDICAL DEVICE notice to its consignees on 03/21/2025 distributed via traceable means. The notice explained the safety issue with the device, potential risk, and requested the following actions be taken: The anesthesia system may continue to be used following the instructions below: Perform the Ventilation Screening Test (specified in Appendix 1) for each affected Carestation system. If the Carestation system passes the Ventilation Screening Test, you can continue to use the device in accordance with the instructions in the User Reference Manual (URM). If the Carestation system fails the Ventilation Screening Test and must be used prior to the system being corrected by GE HealthCare, follow the instructions below: 1. Use the device in accordance with the instructions in the User Reference Manual (URM) with these changes: i. Use only Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) mode to mechanically ventilate a patient. ii. Do not use Volume Control Ventilation (VCV) mode to mechanically ventilate a patient. NOTE: Manual mode of the anesthesia system can be used to provide manual ventilation or allow spontaneous ventilation of the patient. 2. Ensure users are made aware not to use Volume Control Ventilation (VCV) mode on the device until the device has been corrected by GE HealthCare. Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions, and retain this document for your record. GE HealthCare will correct all devices that fail the Ventilation Screening Test at no cost. A GE HealthCare representative will contact you to arrange for the correction. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. Notice sent on 04/21/25 with updated ventilation screening test instructions.

DistributionShow details

Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2624-2026

Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…

  • Pulmonary Valve Prosthesis Percutaneously Delivered
  • Process control
Medtronic Heart Valves…CAJul 8, 2026
Class IOngoingZ-2622-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026
Class IOngoingZ-2621-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026