Skip to content
RecallWatchMedical Device Safety
Class IOngoingZ-1680-2025

Bausch & Lomb Surgical, Inc. recalls Intraocular Lens. enVista Envy IOL

Bausch & Lomb Surgical, Inc.Clearwater, FL, United StatesReported May 14, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Intraocular Lens. enVista Envy IOL, All models starting with EN; enVista Envy Toric IOL, All models starting with ETN.

Lot / code information

UDI
enVista Envy IOL (ENU)/3Q24764, 3Q24935, 3Q25113, 3Q25833, 3Q26256, 3Q26257 — +1543 moreShow all
enVista Envy IOL (ENU)/3Q24764, 3Q24935, 3Q25113, 3Q25833, 3Q26256, 3Q26257, 3Q26538, 3Q26601, 3Q26603, 3Q26604, 3Q26679, 3Q26680, 3Q27353, 3Q27447, 3Q27457, 3Q27905, 3Q27907, 3Q28001, 3Q28002, 3Q28124, 3Q28125, 3Q28801, 3Q28901, 3Q28903, 3Q28904, 3Q28905, 3Q28906, 3Q29020, 3Q29021, 3Q29205, 3Q29266, 3Q29301, 3Q29302, 3Q29304, 3Q29305, 3Q29306, 3Q29401, 3Q29403, 3Q29405, 3Q29406, 3Q29545, 3Q29547, 3Q29548, 3Q29636, 3Q29637, 3Q29835, 3Q29837, 3Q30210, 3Q30310, 3Q30311, 3Q33702, 3Q33822, 3Q33918, 3Q33921, 3Q33924, 3Q33925, 3Q33928, 3Q34031, 3Q34034, 3Q34205, 3Q35006, 3Q35733, 3Q35735, 3Q35736, 3Q35738, 3Q35813, 3R01610, 3R01611, 3R02532, 3R02534, 3R02536, 3R02537, 3R02540, 3R02542, 3R02551, 3R02553, 3R02608, 3R02749, 3R02750, 3R02846, 3R03533, 3R03534, 3R03536, 3R03804, 3R03805, 3R03806, 3R04414, 3R04533, 3R05051, 3R05411, 3Q24749, 3Q24750, 3Q24751, 3Q24756, 3Q24763, 3Q24765, 3Q24766, 3Q24837, 3Q24842, 3Q24934, 3Q25115, 3Q25116, 3Q25117, 3Q25118, 3Q25119, 3Q25421, 3Q25422, 3Q25423, 3Q25424, 3Q25506, 3Q25834, 3Q25922, 3Q25923, 3Q26525, 3Q26527, 3Q26539, 3Q26602, 3Q26605, 3Q27252, 3Q27321, 3Q27322, 3Q27323, 3Q27324, 3Q27354, 3Q27456, 3Q27458, 3Q27906, 3Q27908, 3Q28003, 3Q28004, 3Q28011, 3Q28012, 3Q28802, 3Q29017, 3Q29018, 3Q29022, 3Q29203, 3Q29204, 3Q29303, 3Q29402, 3Q29404, 3Q29546, 3Q29550, 3Q29635, 3Q30211, 3Q30213, 3Q32503, 3Q33816, 3Q33817, 3Q33818, 3Q33819, 3Q33820, 3Q33821, 3Q33823, 3Q33920, 3Q33922, 3Q33926, 3Q33927, 3Q33929, 3Q34032, 3Q34033, 3Q34201, 3Q34202, 3Q34203, 3Q34204, 3Q34901, 3Q34903, 3Q35028, 3Q35031, 3Q35734, 3Q35812, 3R01612, 3R02136, 3R02535, 3R02539, 3R02541, 3R02543, 3R02601, 3R02602, 3R02603, 3R02607, 3R02710, 3R02711, 3R02845, 3R02848, 3R03411, 3R03502, 3R03643, 3R04411, 3R04412, 3R04531, 3R05048, 3R05049, 3R05052, 3R05627/ 10757770636478, 10757770637017, 10757770637031, 10757770637055, 10757770637079, 10757770637543, 10757770637093, 10757770637109, 10757770637536, 10757770637116, 10757770637123, 10757770637130, 10757770637147, 10757770637154, 10757770637161, 10757770637178, 10757770637185, 10757770637192, 10757770637208, 10757770637215, 10757770637222, 10757770637239, 10757770637246, 10757770637253, 10757770636485, 10757770637550, 10757770637277, 10757770637284, 10757770637291, 10757770637307, 10757770637314, 10757770637321, 10757770637338, 10757770637345, 10757770637352, 10757770637369, 10757770637376, 10757770637383, 10757770637390, 10757770637406, 10757770637413, 10757770637420, 10757770637437, 10757770637444, 10757770637451, 10757770637468, 10757770637475, 10757770637482, 10757770637499, 10757770637505, 10757770637512, 10757770637529, 10757770636492; enVista Envy Toric IOL (ETNU)/3Q25926, 3Q25930, 3Q26240, 3Q26426, 3Q26431, 3Q26435, 3Q26436, 3Q26440, 3Q26441, 3Q26621, 3Q26726, 3Q26727, 3Q26731, 3Q26819, 3Q26820, 3Q26910, 3Q26928, 3Q27216, 3Q27217, 3Q27219, 3Q27221, 3Q27223, 3Q27249, 3Q27250, 3Q27251, 3Q27264, 3Q27327, 3Q27330, 3Q27505, 3Q27506, 3Q27518, 3Q27742, 3Q27912, 3Q28007, 3Q28008, 3Q28009, 3Q28010, 3Q28013, 3Q28134, 3Q28136, 3Q28721, 3Q28809, 3Q28818, 3Q28908, 3Q28909, 3Q28910, 3Q29208, 3Q29210, 3Q29212, 3Q29409, 3Q29517, 3Q29642, 3Q29922, 3Q29959, 3Q29961, 3Q30005, 3Q30136, 3Q30137, 3Q30138, 3Q30139, 3Q30208, 3Q30309, 3Q30453, 3Q30454, 3Q30517, 3Q30646, 3Q30647, 3Q30725, 3Q30741, 3Q30744, 3Q30803, 3Q31102, 3Q31107, 3Q31108, 3Q31509, 3Q31540, 3Q31541, 3Q31543, 3Q31865, 3Q31866, 3Q31929, 3Q32003, 3Q32004, 3Q32005, 3Q32009, 3Q32010, 3Q32206, 3Q32209, 3Q32211, 3Q32330, 3Q32331, 3Q32439, 3Q32443, 3Q32525, 3Q32529, 3Q32530, 3Q32606, 3Q32735, 3Q32737, 3Q32743, 3Q32842, 3Q32844, 3Q32846, 3Q32851, 3Q32931, 3Q32932, 3Q33014, 3Q33015, 3Q33128, 3Q33203, 3Q33505, 3Q33508, 3Q33545, 3Q33604, 3Q33613, 3Q33721, 3Q33902, 3Q34121, 3Q34256, 3Q34258, 3Q34315, 3Q35117, 3Q35335, 3Q35430, 3Q35431, 3Q35432, 3Q35433, 3Q35532, 3Q35557, 3Q35624, 3Q35625, 3Q35626, 3Q35628, 3Q35631, 3Q35704, 3Q35741, 3Q35758, 3Q35942, 3R00250, 3R01019, 3R01604, 3R01905, 3R02801, 3R03722, 3R03807, 3R04116, 3Q24946, 3Q25501, 3Q25605, 3Q25606, 3Q25703, 3Q25704, 3Q25722, 3Q25723, 3Q25725, 3Q25740, 3Q25741, 3Q25835, 3Q25836, 3Q25838, 3Q25839, 3Q25924, 3Q25925, 3Q25928, 3Q25931, 3Q25932, 3Q25933, 3Q26003, 3Q26004, 3Q26011, 3Q26012, 3Q26013, 3Q26059, 3Q26060, 3Q26061, 3Q26062, 3Q26063, 3Q26075, 3Q26076, 3Q26077, 3Q26078, 3Q26079, 3Q26080, 3Q26101, 3Q26102, 3Q26211, 3Q26212, 3Q26213, 3Q26214, 3Q26215, 3Q26216, 3Q26217, 3Q26230, 3Q26231, 3Q26232, 3Q26233, 3Q26234, 3Q26235, 3Q26236, 3Q26237, 3Q26238, 3Q26239, 3Q26241, 3Q26259, 3Q26260, 3Q26320, 3Q26321, 3Q26322, 3Q26323, 3Q26324, 3Q26325, 3Q26326, 3Q26327, 3Q26427, 3Q26428, 3Q26430, 3Q26432, 3Q26433, 3Q26434, 3Q26437, 3Q26439, 3Q26442, 3Q26443, 3Q26523, 3Q26524, 3Q26620, 3Q26623, 3Q26624, 3Q26625, 3Q26626, 3Q26721, 3Q26722, 3Q26723, 3Q26724, 3Q26725, 3Q26728, 3Q26729, 3Q26730, 3Q26732, 3Q26821, 3Q26822, 3Q26823, 3Q26824, 3Q26901, 3Q26902, 3Q26903, 3Q26904, 3Q26905, 3Q26908, 3Q26909, 3Q26930, 3Q27218, 3Q27220, 3Q27222, 3Q27224, 3Q27225, 3Q27248, 3Q27256, 3Q27265, 3Q27325, 3Q27326, 3Q27328, 3Q27329, 3Q27331, 3Q27332, 3Q27333, 3Q27334, 3Q27335, 3Q27350, 3Q27351, 3Q27352, 3Q27449, 3Q27501, 3Q27502, 3Q27503, 3Q27504, 3Q27516, 3Q27517, 3Q27519, 3Q27624, 3Q27625, 3Q27626, 3Q27627, 3Q27628, 3Q27629, 3Q27743, 3Q27744, 3Q27745, 3Q27909, 3Q27910, 3Q27911, 3Q27913, 3Q27914, 3Q28005, 3Q28006, 3Q28014, 3Q28127, 3Q28129, 3Q28130, 3Q28131, 3Q28132, 3Q28133, 3Q28135, 3Q28228, 3Q28719, 3Q28806, 3Q28807, 3Q28808, 3Q28810, 3Q28907, 3Q28911, 3Q29104, 3Q29105, 3Q29106, 3Q29128, 3Q29129, 3Q29130, 3Q29206, 3Q29207, 3Q29209, 3Q29211, 3Q29213, 3Q29214, 3Q29307, 3Q29308, 3Q29309, 3Q29310, 3Q29311, 3Q29312, 3Q29407, 3Q29408, 3Q29410, 3Q29411, 3Q29412, 3Q29451, 3Q29452, 3Q29515, 3Q29516, 3Q29518, 3Q29519, 3Q29520, 3Q29522, 3Q29523, 3Q29643, 3Q29644, 3Q29645, 3Q29705, 3Q29706, 3Q29707, 3Q29708, 3Q29709, 3Q29710, 3Q29711, 3Q29712, 3Q29713, 3Q29801, 3Q29838, 3Q29920, 3Q29921, 3Q29923, 3Q29957, 3Q29958, 3Q29960, 3Q30001, 3Q30002, 3Q30003, 3Q30004, 3Q30140, 3Q30201, 3Q30202, 3Q30203, 3Q3

What the firm is doing

On 3/27/2025, a news release was published informing customers that the firm will send out detailed instructions for what to do with any stock on hand. On 4/4/2025, the news release was updated and customers were also informed in a recall notice of the following: For Patients: Cataract surgery patients experiencing eye discomfort should contact their eyecare professional immediately. For Eyecare Professionals: Continue to closely monitor cataract surgery patients for at least 48 hours after surgery. Please instruct patients with new or worsening symptoms to contact your office immediately. Customers were asked to do the following: 1) Quarantine and do not use affected products. 2) Complete and return the acknowledgment form with returned devices or if there are no affected devices to return email the form to BLSurgCustSvc@Bausch.com 3) Provide this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred (as appropriate). Please transfer this notice to other organizations on which this action has an impact (as appropriate). Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. 4) All reported cases of TASS to date have occurred within 12 - 48 hours after eye surgery. Rare cases of TASS could have delayed onset, so we request post-operative patient care protocols be followed to closely monitor patients who have received an intraocular lens impacted by this recall. Firm is investigating to identify the root cause of the issue. If you have any questions, or to make a report, contact our Customer Care team at (800) 338-2020, option 9 or 1. On 5/1/2025, revised recall notices were sent informing customers that the firm is reducing the recall scope.

DistributionShow details

Worldwide distribution. US Nationwide including PR and GU, and the countries of CAN, AUS, AUT, BEL, BGR, IC, HRV, CZE, FRA, DEU, GRC, HUN, ITA, NLD, POL, PRT, ROU, ESP, GBR, NZL, HKG.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2879-2026

Boston Scientific Neuromodulation Corporation recalls INFINION CX LEAD KIT

Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, i…

  • Totally Implanted For Pain Relief Spinal-Cord Stimulator
  • Device Design
Boston Scientific Neur…CAAug 12, 2026