Class IOngoingZ-1681-2025

Smiths Medical ASD, Inc. recalls CADD-Solis Infusion Pump Models: 1) PUMP KIT

Smiths Medical ASD, Inc.Minneapolis, MN, United StatesReported May 14, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

CADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-51; 2) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-51; 3) PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA, Model Number: 21-2112-0100-51; 4) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2112-0300-01; 5) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2111-0300-01; 6) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-51; 7) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 8) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-51; 9) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 10) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, CE ENGLISH 1/EA, Model Number: 21-2111-0400-51; 11) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2111-0300-50; 12) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 13) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 14) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-78; 15) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 16) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2112-0300-00; 17) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-78; 18) PUMP, DEMO, CADD-SOLIS, MODEL 2100, NOT FOR HUMAN USE 1/EA, Model Number: 21-2101-249; 19) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-78; 20) PUMP KIT, CADD-SOLIS, YELLOW, MDL 2100 1/EA, Model Number: 21-2102-51; 21) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, NFHU, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-249; 22) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA, Model Number: 21-2111-0300-231; 23) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-78; 24) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0400-51; 25) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2112-0300-50; 26) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 27) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 28) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 29) PUMP, CADD-SOLIS, MODEL 2110, CE ENGLISH OUS 1/EA, Model Number: 21-2111-0100-50; 30) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 31) PUMP KIT, CADD-SOLIS HPCA, YELLOW KEYPAD, MODEL 2110, CE ENGLISH, OUS 1/EA, Model Number: 21-2112-0100-50; 32) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 33) PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA, Model Number: 21-2112-0400-01.

Lot / code information

Serial #
. 1 ) 21-2111-0402-51
UDI
15019517154290 ; 2 ) 21-2112-0402-51
UDI
15019517154993 ; 3 ) 21-2112-0100-51
UDI
10610586039232 ; 7 ) 21-2111-0100-51
UDI
30610586039205 ; 9 ) 21-2101-51
UDI
10610586031854 ; 13 ) 21-2111-0300-00
UDI
10610586038792 ; 14 ) 21-2111-0402-78
UDI
15019517154887 ; 15 ) 21-2111-0100-00
UDI
10610586039195 ; 16 ) 21-2112-0300-00
UDI
10610586038822 ; 17 ) 21-2112-0402-78
Show 15 more code fields
UDI
15019517155013 ; 18 ) 21-2101-249
UDI
10610586034787 ; 20 ) 21-2102-51
UDI
10610586033933 ; 22 ) 21-2111-0300-231
UDI
10610586038785 ; 23 ) 21-2112-0401-78
UDI
15019517101034 ; 24 ) 21-2112-0400-51
UDI
10610586041051 ; 25 ) 21-2112-0300-50
UDI
10610586038860 ; 26 ) 21-2111-0300-00
UDI
10610586038792 ; 27 ) 21-2111-0100-00
UDI
10610586039195 ; 28 ) 21-2111-0100-51
UDI
30610586039205 ; 29 ) 21-2111-0100-50
UDI
10610586039256 ; 30 ) 21-2101-51
UDI
10610586031854 ; 31 ) 21-2112-0100-50
UDI
10610586039270 ; 32 ) 21-2101-00
UDI
35019517154904; 33 ) 21-2112-0400-01
UDI
10610586041822

What the firm is doing

An URGENT MEDICAL DEVICE CORRECTION dated 4/10/25 was sent to customers. Actions for Users: Inform all affected CADD-Solis and CADD-Solis VIP users of this notice. Provide the instructions below: 1. Ensure all users or potential users of these products are immediately made aware of this notification. 2. Upon receipt of this notification and ongoing during regular maintenance, visually inspect the external condition of the battery pack or Wireless Communication Module and look for evidence of any physical damage. Handle a damaged device with caution. As stated in the battery pack Instructions for Use, if the battery pack housing is cracked or otherwise damaged, replace the battery pack. NEVER use a battery pack that appears damaged. A rechargeable battery pack must be replaced with either another CADD-Solis rechargeable battery pack or with 4 AA batteries. 3. Visually inspect the condition of the battery compartment for evidence of any physical damage. Contact Technical Support for repairs. 4. Always check the battery compartment for fluid or debris before inserting the batteries, and do not allow any fluid or debris to fall into the battery compartment. 5. Complete the electronic Response Form via the link provided within this email notification within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product. 6. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward your customer list to Sedgwick at icumedical3393@sedgwick.com. Sedgwick will notify customers with email addresses via email and will send physical notifications to customers without email addresses in order for them to complete the Response Form Follow up Actions by Smiths Medical: Smiths Medical is continuing to investigate this matter to determine if additional actions may be warranted. For further inquiries, please contact Smiths Medical using the following information: Smit

DistributionShow details

Worldwide - US Nationwide distribution.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2624-2026

Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…

  • Pulmonary Valve Prosthesis Percutaneously Delivered
  • Process control
Medtronic Heart Valves…CAJul 8, 2026
Class IOngoingZ-2622-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination