Getinge Disinfection Ab recalls Washer Disinfector Aquadis 56
Reason for recall
The frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Washer Disinfector Aquadis 56; Model Number: 56M
Lot / code information
- Model
- 56M
- UDI
- 07340153710351
- Serial #
WAA105199, WAA105251, WAA105268, WAA105357, WAA105535, WAA105700 — +418 moreShow all
WAA105199, WAA105251, WAA105268, WAA105357, WAA105535, WAA105700, WAA105915, WAA105972, WAA105973, WAA106079, WAA106114, WAA106135, WAA106451, WAA106471, WAA106487, WAA106505, WAA107436, WAA107492, WAA107565, WAA107581, WAA107588, WAA107920, WAA108225, WAA108807, WAA108920, WAA108921, WAA108922, WAA109036, WAA109047, WAA109231, WAA109377, WAA109392, WAA109635, WAA109687, WAA109731, WAA109790, WAA109835, WAA109840, WAA109853, WAA109907, WAA109972, WAA110109, WAA110366, WAA110487, WAA110546, WAA110557, WAA110711, WAA110726, WAA110802, WAA110847, WAA110862, WAA110984, WAA110999, WAA111007, WAA111011, WAA111068, WAA111086, WAA111087, WAA111200, WAA600033, WAA600054, WAA600100, WAA600102, WAA600103, WAA600145, WAA600190, WAA600282, WAA600369, WAA600431, WAA600434, WAA600444, WAA600445, WAA600553, WAA600554, WAA600587, WAA600588, WAA600609, WAA600610, WAA600704, WAA106987, WAA110250, WAA110257, WAA108634, WAA110823, WAA108206, WAA109799, WAA109066, WAA109076, WAA109091, WAA109373, WAA110528, WAA111028, WAA600134, WAA600189, WAA110820, WAA110123, WAA108871, WAA108872, WAA106724, WAA109403, WAA109405, WAA109495, WAA109933, WAA110513, WAA110527, WAA110617, WAA109658, WAA109672, WAA110029, WAA110779, WAA110893, WAA107614, WAA108608, WAA108610, WAA109222, WAA109764, WAA109782, WAA109946, WAA110055, WAA110083, WAA110204, WAA110801, WAA600066, WAA600068, WAA600072, WAA600507, WAA600508, WAA106520, WAA106618, WAA106639, WAA110594, WAA110595, WAA110842, WAA111150, WAA600105, WAA108702, WAA108040, WAA106692, WAA107486, WAA107491, WAA108832, WAA108998, WAA109241, WAA109244, WAA109885, WAA110167, WAA110279, WAA110290, WAA110303, WAA110349, WAA110394, WAA110406, WAA111093, WAA111098, WAA111112, WAA600216, WAA600441, WAA600509, WAA600560, WAA600611, WAA600612, WAA600613, WAA600648, WAA107279, WAA107940, WAA109140, WAA109973, WAA110534, WAA110545, WAA111142, WAA111174, WAA111177, WAA111185, WAA106240, WAA106760, WAA110566, WAA110876, WAA108501, WAA108502, WAA108514, WAA109093, WAA109409, WAA109805, WAA109808, WAA109966, WAA110032, WAA110037, WAA110618, WAA600141, WAA600220, WAA600221, WAA600222, WAA600223, WAA600253, WAA600254, WAA600255, WAA600450, WAA600451, WAA600585, WAA600586, WAA110881, WAA105641, WAA105682, WAA106980, WAA107125, WAA108151, WAA108367, WAA108394, WAA108519, WAA108531, WAA108953, WAA109653, WAA109874, WAA110018, WAA110027, WAA110056, WAA110327, WAA111114, WAA600106, WAA600186, WAA600187, WAA107148, WAA107232, WAA107267, WAA107307, WAA107360, WAA107418, WAA108067, WAA108136, WAA108137, WAA108161, WAA108170, WAA108217, WAA108409, WAA108479, WAA108499, WAA108500, WAA108653, WAA108743, WAA108745, WAA108746, WAA108863, WAA108864, WAA109187, WAA109199, WAA109200, WAA109207, WAA109269, WAA109278, WAA109285, WAA109305, WAA109309, WAA109314, WAA109327, WAA109497, WAA109501, WAA109507, WAA109565, WAA110208, WAA110226, WAA110361, WAA110643, WAA110836, WAA110845, WAA110937, WAA110938, WAA111012, WAA111032, WAA111055, WAA600047, WAA600053, WAA600135, WAA600256, WAA600286, WAA600287, WAA600288, WAA600370, WAA600371, WAA600436, WAA600438, WAA600442, WAA600446, WAA600448, WAA600555, WAA110297, WAA110315, WAA107182, WAA109948, WAA110950, WAA110268, WAA110648, WAA110663, WAA107190, WAA107199, WAA109370, WAA109513, WAA110486, WAA106247, WAA106399, WAA106412, WAA108983, WAA110013, WAA110420, WAA600217, WAA600649, WAA600858, WAA109765, WAA109696, WAA109713, WAA109737, WAA109182, WAA109673, WAA109686, WAA109862, WAA109864, WAA110256, WAA110274, WAA110460, WAA109949, WAA106546, WAA106548, WAA107084, WAA107905, WAA108894, WAA109343, WAA109352, WAA109393, WAA109908, WAA109931, WAA110686, WAA110833, WAA111083, WAA600218, WAA600219, WAA600278, WAA600556, WAA107522, WAA107594, WAA108973, WAA108975, WAA110638, WAA111052, WAA111053, WAA111186, WAA111189, WAA110395, WAA107136, WAA110323, WAA110338, WAA110360, WAA110370, WAA107215, WAA106992, WAA108059, WAA108623, WAA110703, WAA105696, WAA107624, WAA107641, WAA107691, WAA107699, WAA107731, WAA107736, WAA107746, WAA107756, WAA107763, WAA107788, WAA107790, WAA107801, WAA107804, WAA107806, WAA107825, WAA107829, WAA107858, WAA107860, WAA107883, WAA107897, WAA108226, WAA108264, WAA108267, WAA108300, WAA108301, WAA108410, WAA108466, WAA108952, WAA109249, WAA109260, WAA109261, WAA109266, WAA109340, WAA109688, WAA109757, WAA109937, WAA110374, WAA110428, WAA110432, WAA110452, WAA110453, WAA600011, WAA600082, WAA600101, WAA600104, WAA600107, WAA600129, WAA600130, WAA600285, WAA600551, WAA600552, WAA600559, WAA109920, WAA109929, WAA600280, WAA600281, WAA110756, WAA110757, WAA109631, WAA109916, WAA110235, WAA110248, WAA600289, WAA600327, WAA110154, WAA600650, WAA107607
What the firm is doing
On April 2, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions 1. Check the serial number of yourWasher Disinfector Aquadis 56 to determine if it is affected. The serial number can be found on the marking label, located on the inside of the front Panel door. 2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 3. Complete and return the enclosed response form Getinge by e-mailing a copy to recallresponses.qrc@getinge.com. Note: If your unit is covered by a service agreement, Getinge will carry out the testing and required rework, if necessary, during the next regularly scheduled maintenance. If your facility currently does not have a service agreement with Getinge, our field service technicians will contact you to set up testing. Getinge will conduct testing to determine if the frequency inverter is programmed or not, and if it is found to be affected by this issue, the memory module in the frequency inverter will be replaced, free of charge. Actions to be taken by Getinge: Getinge will conduct testing to determine if the frequency inverter is programmed or not, and if it is found to be affected by this issue, the memory module in the frequency inverter will be replaced, free of charge. This will be performed during your routine planned maintenance to avoid any inconvenience. Additional Information: Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax using the following: - Online: www.accessdata.fda.gov/scripts/medwatch/ - Regular Mail: Download form at www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form - Fax: 1-800-FDA-0178 (1-800-332-0178) We deeply regret this inconve
DistributionShow detailsHide
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lithuania, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, Nigeria, Norway, Oman, Pakistan, Philippines, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Vietnam, Zambia.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1721-2025
- FDA device classification · MECOfficial FDA classification for this device type
- CFR regulation · 21 CFR 880.6992The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Getinge Disinfection AbSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm