Baxter Healthcare Corporation recalls Baxter Novum IQ LVP INFUSION SYSTEM
Reason for recall
There is a potential for the Novum IQ LVP due to the potential for underinfusion following use of the "standby mode" feature or if the device is powered off with the set loaded.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Baxter Novum IQ LVP INFUSION SYSTEM, REF 40700BAXUS
Lot / code information
All serial numbers
What the firm is doing
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 04/24/2025 via USPS First Class Mail. The notice explained the issue, hazard, and requested the following: Actions to be Taken by Customers 1. For flow rates greater than 50 mL/hour, do not exceed a programmed standby time of 2 hours and 30 minutes. Monitor patients frequently to ensure that the appropriate infusion is being delivered. 2. Please remove the set upon powering off the device. 3. Post the enclosed informational poster with Novum IQ LVPs in your facility. 4. Acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. Please forward a copy of this communication to the Chief Medical Officer, Medical Director, Director of Nursing, Director of Pharmacy, Facility Risk Manager, Director of Purchasing/Central Supply, and any other departments within your institution who use the affected product. For general questions regarding this communication, or if you experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867 (select option 2, then option 2 again) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.
DistributionShow detailsHide
US and Canada
A reasonable probability that use of the product will cause serious adverse health consequences or death.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1767-2025
- FDA 510(k) clearance · K211122The device's official FDA premarket clearance record
- FDA device classification · FRNOfficial FDA classification for this device type
- CFR regulation · 21 CFR 880.5725The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Baxter Healthcare CorporationSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class I recalls
Spectra Medical Devices, LLC recalls Brand Name: Elevaris/Spectra Medical Devices Product Name: Sodium Chloride Injection
Manufacturing firm, Huons, was cited for issues regarding product quality and sterility assurance. If the product sterility is compromised, patients may be exposed to microorganisms, endotoxins, or foreign matter.…
- Vascular Access Flush Saline
- Nonconforming Material/Component
Baxter Healthcare Corporation recalls Baxter Duo-Vent Solution Set and Male Luer Lock Adapter
Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position.
- Intravascular Administration Set
- Under Investigation by firm
Boston Scientific Neuromodulation Corporation recalls INFINION CX LEAD KIT
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, i…
- Totally Implanted For Pain Relief Spinal-Cord Stimulator
- Device Design