GE Medical Systems, LLC recalls OEC 3D
Reason for recall
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- OEC 3DGTIN 00001234567895400 affected lotsFAHXTX00002FAHXTX00005FAHXTX00014FAHXTX00018FAHXTX00019FAHXTX00020FAHXTX00024FAHXTX00028
+392 more
FAHXTX00034FAHXTX00037FAHXTX00043FAHXTX00052FAHXTX00063FAHXTX00066FAHXTX00068FAHXTX00070FAHXTX00075FAHXTX00078FAHXTX00079FAHXTX00080FAHXTX00081FAHXTX00083FAHXTX00092FAHXTX00093FAHXTX00097FAHXTX00110FAHXTX00111FAHXTX00112FAHXTX00114FAHXTX00115FAHXTX00116FAHXTX00117FAHXTX00118FAHXTX00119FAHXTX00120FAHXTX00121FAHXTX00122FAHXTX00124FAHXTX00125FAHXTX00135FAHXTX00137FAHXTX00138FAHXTX00151FAHXTX00152FAHXTX00155FAHXTX00156FAHXTX00162FAHXTX00163FAHXTX00166FAHXTX00170FAHXTX00171FAHXTX00177FAHXTX00178FAHXTX00188FAHXTX00192FAHXTX00193FAHXTX00202FAHXTX00204FAHXTX00211FAHXTX00212FAHXTX00217FAHXTX00219FAHXTX00220FAHXTX00221FAHXTX00222FAHXTX00223FAHXTX00224FAHXTX00226FAHXTX00227FAHXTX00229FAHXTX00230FAHXTX00236FAHXTX00240FAHXTX00244FAHXTX00261FAHXTX00265FAHXTX00266FAHXTX00269FAHXTX00271FAHXTX00272FAHXTX00273FAHXTX00274FAHXTX00280FAHXTX00283FAHXTX00284FAHXTX00296FAHXTX00299FAHXTX00300FAHXTX00304FAHXTX00305FAHXTX00308FAHXTX00312FAHXTX00313FAHXTX00320FAHXTX00323FAHXTX00325FAHXTX00332FAHXTX00340FAHXTX00342FAHXTX00343FAHXTX00352FAHXTX00353FAHXTX00354FAHXTX00357FAHXTX00358FAHXTX00359FAHXTX00360FAHXTX00363FAHXTX00376FAHXTX00379FAHXTX00381FAHXTX00382FAHXTX00390FAHXTX00392FAHXTX00399FAHXTX00400FAHXTX00410FAHXTX00413FAHXTX00414FAHXTX00415FAHXTX00420FAHXTX00421FAHXTX00429FAHXTX00439FAHXTX00453FAHXTX00455FAHXTX00456FAHXTX00463FAHXTX00465FAHXTX00466FAHXTX00479FAHXTX00481FAHXTX00483FAHXTX00484FAHXTX00486FAHXTX00488FAHXTX00495FAHXTX00497FAHXTX00498FAHXTX00507FAHXTX00509FAHXTX00513FAHXTX00523FAHXTX00530FAHXTX00534FAHXTX00538FAHXTX00539FAHXTX00540FAHXTX00547FAHXTX00549FAHXTX00555FAHXTX00557FAHXTX00560FAHXTX00565FAHXTX00569FAHXTX00570FAHXTX00571FAHXTX00572FAHXTX00575FAHXTX00576FAHXTX00579FAHXTX00592FAHXTX00595FAHXTX00610FAHXTX00611FAHXXX00029FAHXXX00030FAHXXX00038FAHXXX00040FAHXXX00056FAHXXX00057FAHXXX00059FAHXXX00060FAHXXX00061FAHXXX00069FAHXXX00082FAHXXX00085FAHXXX00086FAHXXX00089FAHXXX00091FAHXXX00093FAHXXX00097FAHXXX00098FAHXXX00099FAHXXX00105FAHXXX00106FAHXXX00112FAHXXX00113FAHXXX00114FAHXXX00116FAHXXX00117FAHXXX00119FAHXXX00123FAHXXX00142FAHXXX00144FAHXXX00145FAHXXX00147FAHXXX00150FAHXXX00151FAHXXX00153FAHXXX00155FAHXXX00164FAHXXX00173FAHXXX00176FAHXXX00178FAHXXX00179FAHXXX00188FAHXXX00190FAHXXX00193FAHXXX00196FAHXXX00199FAHXXX00200FAHXXX00207FAHXXX00208FAHXXX00209FAMHTX00001FAMHTX00005FAMHTX00007FAMHTX00012FAMHTX00014FAMHTX00041FAMHTX00051FAMHTX00055FAMHTX00064FAMHTX00069FAMHTX00081FAMHTX00085FAMHXX00003FAMHXX00011FAMHXX00019FAMHXX00023FAXXTE00003FAXXTE00006FAXXTE00009FAXXTE00011FAXXTE00013FAXXTE00014FAXXTE00015FAXXTE00017FAXXTE00021FAXXTE00022FAXXTE00024FAXXTE00025FAXXTE00028FAXXTE00031FAXXTE00032FAXXTE00033FAXXTE00034FAXXTE00035FAXXTE00037FAXXTE00038FAXXTE00039FAXXTE00040FAXXTE00044FAXXTE00046FAXXTE00047FAXXTE00049FAXXTE00050FAXXTE00051FAXXTE00052FAXXTE00056FAXXTE00062FAXXTE00063FAXXTE00064FAXXTE00067FAXXTE00071FAXXTE00072FAXXTE00073FAXXTE00076FAXXTE00084FAXXTE00085FAXXTE00091FAXXTE00092FAXXTE00094FAXXTE00095FAXXTE00099FAXXTE00100FAXXTE00101FAXXTE00104FAXXTE00105FAXXTE00106FAXXTE00107FAXXTE00109FAXXTE00110FAXXTE00114FAXXTE00116FAXXTE00117FAXXTE00120FAXXTE00121FAXXTE00128FAXXTE00129FAXXTE00134FAXXTE00136FAXXTE00137FAXXTE00141FAXXTE00144FAXXTE00145FAXXTE00147FAXXTE00149FAXXTE00163FAXXTE00164FAXXTE00170FAXXTE00173FAXXTE00175FAXXTE00179FAXXTE00180FAXXTE00182FAXXTE00184FAXXTE00185FAXXTE00195FAXXTE00197FAXXTE00205FAXXTE00212FAXXTE00220FAXXTE00224FAXXTE00225FAXXTE00227FAXXTE00228FAXXTE00229FAXXTE00230FAXXTE00231FAXXTE00234FAXXTE00236FAXXTE00237FAXXTE00238FAXXTE00240FAXXTE00241FAXXTE00243FAXXTE00246FAXXTE00248FAXXTE00253FAXXTE00256FAXXTE00257FAXXTE00258FAXXTE00259FAXXTE00263FAXXTE00267FAXXTE00269FAXXTE00270FAXXTE00271FAXXTE00272FAXXTE00277FAXXTE00278FAXXTE00279FAXXTE00280FAXXTE00282FAXXTE00286FAXXTE00291FAXXTE00293FAXXTE00295FAXXTE00301FAXXTE00317FAXXTE00319FAXXTE00320FAXXTE00321FAXXTE00323FAXXTE00328FAXXTE00329FAXXTE00330FAXXTE00331FAXXTE00334FAXXTE00336FAXXTE00337FAXXTE00338FAXXTE00339FAXXTE00340FAXXTE00342FAXXTE00344FAXXTE00352FAXXTE00353FAXXTE00354FAXXTE00357FAXXTE00358FAXXTE00361FAXXTE00362FAXXTE00363FAXXTE00372FAXXTE00373FAXXTE00376FAXXTE00379FAXXTE00380FAXXTE00383FAXXTE00389FAXXTE00392FAXXTE00393FAXXTE00396FAXXTE00407FAXXTE00413FAXXTE00418FAXXTE00422FAXXTE00424FAXXTE00425FAXXTE00429FAXXTE00431FAXXTE00433FAXXTE00434FAXXTE00436FAXXTE00439FAXXTE00442FAXXTE00443
What the firm is doing
All impacted customers with receive a customer letter notifying them of the potential issue. GE HealthCare will send a field service engineer to perform measurements of the Xray field on the identified systems (listed in Appendix A of the customer letter).
DistributionShow detailsHide
US and OUS
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1820-2025
- FDA 510(k) clearance · K203346The device's official FDA premarket clearance record
- FDA 510(k) clearance · K233669The device's official FDA premarket clearance record
- FDA device classification · OXOOfficial FDA classification for this device type
- CFR regulation · 21 CFR 892.1650The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find GE Medical Systems, LLCSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) BASIC PACK
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Eye Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) KIT
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Orthopedic Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) CHI STD NEURO CRANI OPT 2
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Neurological Tray
- Nonconforming Material/Component