Olympus Corporation Of The Americas recalls Olympus Thunderbeat II Shears w/Ultrasonic Mode
Reason for recall
Potential for detachment of a distal tip component of the device during use.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 45cm. Model Number:TB2-0545FCUDI-DI 04953170440021.Model Number
What the firm is doing
Olympus notified consignees on about 03/23/2026 via emailed letter. Consignees were instructed to immediately cease use of all THUNDERBEAT II devices, examine inventory for affected product and quarantine any on hand, and complete, return the Reply Form provided and arrange for the return of all affected units in inventory. Consignees were also instructed to notify other users if the affected products were transferred or further distributed.
DistributionShow detailsHide
International distribution in the countries of Australia, Japan, Hong Kong, and Europe.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1854-2026
- FDA 510(k) clearance · K252150The device's official FDA premarket clearance record
- FDA device classification · GEIOfficial FDA classification for this device type
- CFR regulation · 21 CFR 878.4400The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Olympus Corporation of the AmericasSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm