Class IOngoingZ-1868-2025

Zyno Medical LLC recalls Z-800WF Infusion System. Intended to provide intravenous infusion of parenteral fluids

Zyno Medical LLCNatick, MA, United StatesReported Jun 18, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Unreleased software versions were installed on distributed devices without verification or validation.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Z-800WF Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

Lot / code information

Model
Z800WF
UDI
00814371020037
Serial #
500004; 500005; 500006; 500810; 500812; 500813; 501748; 503680; 503681; 503682

What the firm is doing

Consignees received an "URGENT: MEDICAL DEVICE CORRECTION" notification dated 5/7/25 by email. The notification instructs consignees to check their stock for affected devices and to immediately stop use; recalled devices are to be returned to Zyno Medical via the provided Response Verification Form. The recall notice is to be provided to personnel at consignee locations who need to be informed and communicated to facilities to where devices were transferred to. Any questions about this recall action are to be directed via email to Frederick Lee at feedback@intuvie.com. A representative from Zyno Medical s partner company, Intuvie LLC, will be in contact with you to coordinate instructions on exchanging devices to remediate this action.

DistributionShow details

US Nationwide

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2879-2026

Boston Scientific Neuromodulation Corporation recalls INFINION CX LEAD KIT

Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, i…

  • Totally Implanted For Pain Relief Spinal-Cord Stimulator
  • Device Design
Boston Scientific Neur…CAAug 12, 2026