Dexcom, Inc. recalls Dexcom G7 iOS Application
Reason for recall
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose alerts will not trigger, there is potential for the missed detection of a hyperglycemic or hypoglycemic event or a treatment decision made based off incorrect data. There is no impact to transmitter communication with concurrently connected displays. A concurrently connected receiver, Apple Watch, and/or insulin pump will continue to receive EGVs directly from the transmitter when the app is in the problematic state. There is no impact to Automated Insulin Dosing (AID) therapy as the AID system continues to receive EGVs directly from the transmitter.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) systemUDI 00386270001863
What the firm is doing
On May 05, 2025, Dexcom issued a "Urgent Warning" Recall notification to affected consignees via application update. Dexcom asked consignees to take the following actions: 1. Upgrading from app version 2.7 is mandatory and app version 2.7 is no longer available for use. 2. If you need assistance, contact Technical Support. Local Tech Support contact information is available at Dexcom.com Please note that US customers should utilize one of the following options for support. " Phone - 1-844-478-1600 " Call Back request Call Back Request Form " Web Form - https://dexcom.custhelp.com/app/webform " Live chat - https://dexcom.custhelp.com/app/chat/chat_avail If you have any questions, call 858-200-0200 or email: Charles.boykin@dexcom.com.
DistributionShow detailsHide
Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: United Arab Emirates (AE), Austria (AT), Australia (AU), Belgium (BE), Bahrain (BH), Canada (CA), Switzerland (CH), Czech Republic (CZ), Germany (DE), Denmark (DK), Spain (ES), Finland (FI), France (FR), United Kingdom (GB), Hong Kong (HK), Ireland (IE), Israel (IL), Iceland (IS), Italy (IT), Jordan (JO), Korea, Repulic (KR), Kuwait (KW), Lebanon (LB), Lithuania (LT), Luxemburg (LU), Malta (MT), Netherlands (NL), Norway (NO), New Zealand (NZ), Oman (OM), Poland (PL), Portugal (PT), Qatar (QA), Saudi Arabia (SA), Sweden (SE), Slovenia (SI), Slovakia (SK), Turkey (TR) and South Africa (ZA).
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1906-2025
- FDA 510(k) clearance · K231081The device's official FDA premarket clearance record
- FDA device classification · QBJOfficial FDA classification for this device type
- CFR regulation · 21 CFR 862.1355The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Dexcom, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) BASIC PACK
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Eye Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) KIT
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Orthopedic Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) CHI STD NEURO CRANI OPT 2
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Neurological Tray