Class IOngoingZ-1944-2025

Dexcom, Inc. recalls Dexcom G7 Glucose Receiver

Dexcom, Inc.San Diego, CA, United StatesReported Jun 25, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0), STK-GT-001 (MT26403-1), STK-GT-008 (MT26403-5), STK-GT-013 (MT26403-1) , STK-GT-019 (MT26403-4) , STK-GT-100 (MT26403-2), STK-GT-109 (MT26403-2) , STK-GT-113 (MT26403-3)
    UDI STK-AT-011
    20 affected lots
    STK-AT-01100386270002198STK-AT-012-00386270002518STK-AT-01300386270002525STK-GT-00100386270004291
    +12 moreSTK-GT-00800386270004260STK-GT-01300386270002273STK-GT-01904595986293080STK-GT-10000386270002020STK-GT-10900386270002129STK-GT-11300386270003638

What the firm is doing

On May 12, 2025, Dexcom issued a "Urgent: Medical Device Correction" Notification to affected consignees via E-Mail. On June 9, 2025, Dexcom issued an Urgent: Medical Device Correction update Dexcom asked consignees to take the following actions: 1. Verify if your receiver is affected by this issue: Check the serial number for any receiver(s) you use: https://dexcom.com/en-us/checkreceiver 2. If your receiver is affected, contact Dexcom Technical Support at 1-844-478-1600 to coordinate return of the affected receiver and a no charge replacement. 3. Whether or not your receiver is affected by this notification, Dexcom continues to recommend regular testing of the speaker. Test speaker function every time you charge your receiver. When you charge, a reminder prompt will invite you to test speaker function. 4.If you need assistance related to this issue or notification, please call Dexcom Technical Support at 1- 844-478-1600. Technical Support is available 24 hours a day/7 days a week to assist you.

DistributionShow details

worldwide distribution - US Nationwide and the countries of Andorra, Argentina Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, New Zealand, Norway, Oman, Poland, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2879-2026

Boston Scientific Neuromodulation Corporation recalls INFINION CX LEAD KIT

Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, i…

  • Totally Implanted For Pain Relief Spinal-Cord Stimulator
  • Device Design
Boston Scientific Neur…CAAug 12, 2026