Baxter Healthcare Corporation recalls Baxter Novum IQ Syringe Pump
Reason for recall
Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Baxter Novum IQ Syringe Pump, product code 40800BAXUS,
Lot / code information
UDI/DI 05413765852428, All Serial Numbers
What the firm is doing
Baxter issued an URGENT MEDICAL DEVICE RECALL notice to is consignees on 05/12/2025 via USPS First Class Mail. The notice explained the issue, risk, and requested the following: Actions to be Taken by Customers 1. Customers may continue to use the Novum IQ syringe pump until the pumps can be repaired. If the pump detects a syringe when no syringe is present, or you experience System Error 21502 (Flange Sensor Failure), power cycle the device and reload the syringe. Complete the therapy and remove the pump from service. 2. A Baxter representative will contact your facility to coordinate the repair of your Novum IQ syringe pumps. Please note you will be receiving this repair from Baxter at no charge. 3. Acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 4. Please forward a copy of this communication to the Chief Medical Officer, Medical Director, Director of Nursing, Director of Pharmacy, Facility Risk Manager, Director of Purchasing/Central Supply, and any other departments within your institution who use the affected product. For general questions regarding this communication, or if you experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867, Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.
DistributionShow detailsHide
US Nationwide distribution.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-1964-2025
- FDA 510(k) clearance · K242390The device's official FDA premarket clearance record
- FDA device classification · FRNOfficial FDA classification for this device type
- CFR regulation · 21 CFR 880.5725The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Baxter Healthcare CorporationSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm