Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-1974-2025

BD SWITZERLAND SARL recalls BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP

BD SWITZERLAND SARLEysins, SwitzerlandReported Jun 18, 2025 · 1 year ago
Legal News Analyst ·

Reason for recall

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP1 code
  • 10012241
BD TEXIUM (China)1 code
  • 10012241 CHINA

Lot / code information

Lot #
(Expiration): 10012241/07613203010450/23015679(January 31, 2026), 23035092(March 6, 2026), 23055076(May 4, 2026) — +29 moreShow all
(Expiration): 10012241/07613203010450/23015679(January 31, 2026), 23035092(March 6, 2026), 23055076(May 4, 2026), 23075108(July 5, 2026), 23095125(September 8, 2026), 23105028(October 3, 2026), 23125147(December 5, 2026), 24015246(January 11, 2027), 24026422(February 27, 2027), 24036125(March 25, 2027), 24045689(April 24, 2027), 24066894(June 25, 2027), 24075112(July 26, 2027), 24095557(September 24, 2027), 24115004(November 1, 2027); 10012241 CHINA/10885403238383/23095707(September 14, 2026), 24025890(February 7, 2027), 24105002(October 3, 2027)

What the firm is doing

On 5/12/2025, removal notices were mailed or emailed to Medical Director, Risk Manager, Medical Device Safety Officer, Nurse Managers, and Distributors who were asked to do the following: 1. Immediately review your inventory for affected product. Destroy all unused affected products subject to the recall following your institution s process for destruction. 2. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 3. Complete the Customer Response Form via email to bdrc53@bd.com 4. In addition, distributors were asked to identify all customers that purchased any affected product and to provide a copy of the customer letter to all customers and to advise them of the removal notification. Report any complaints experienced with the use of this product to the firm's North American Regional Complaint Center: Phone: 1-844-823-5433 Mon-Fri 8:00am and 5:00pm CT, Email: productcomplaints@bd.com

DistributionShow details

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.