Medline Industries, LP recalls Centurion medical convenience kits
Reason for recall
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Centurion medical convenience kits, containing Webcol Large Alcohol Prep Pads, labeled as: Universal Central/Art Line Insertion Kit, Medline Kit SKUART155A2 affected lots20251008902026011290
What the firm is doing
Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/16/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the affected product be quarantined pending additional labeling. Consignees were directed to respond to the notice via Website link: https://recalls.medline.com, and upon receipt of the submitted response form, the account will receive over-labels to place on affected inventory, with instructions for the user to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight. Distributors or those who have further distributed the product are directed to notify those receiving the affected kits, and to include the customers quantities on the response form. Upon receipt of your submitted response form, you will receive over-labels to provide to the customers to affix to their affected product. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
DistributionShow detailsHide
US Nationwide distribution.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2039-2026
- FDA device classification · CBTOfficial FDA classification for this device type
- CFR regulation · 21 CFR 868.1100The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medline Industries, LPSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm