PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Integris-Allura system
Reason for recall
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (722030), Integris Allura 15 & 12 (monoplane) (722043), INTEGRIS Allura 15-12 (biplane) (722044), Integris-Allura 9 (722018), Integris-Allura 9 (Biplane) (722021), Integris CV Cesar-Powerpack (722030), Integris H5000 C / Allura 9C (722016);
Lot / code information
- Model
- 722030
- UDI
- N/A
- Serial #
- 4888270, 59, 0, 33; 2. Integris Allura 15 & 12 (monoplane) (722043)
- UDI
- N/A
- Serial #
- 365, 18151, 176, 239, 407, 426, 0; 3. INTEGRIS Allura 15-12 (biplane) (722044)
- UDI
- N/A
- Serial #
- 86322, 30, 66; 4. Integris-Allura 9 (722018)
- UDI
- N/A
- Serial #
- 26, 85; 5. Integris-Allura 9 (Biplane) (722021)
- UDI
- N/A
- Serial #
- 22, 75, 25, 108; 6. Integris H5000 C / Allura 9C (722016)
- UDI
- N/A
- Serial #
- 4059001008, 4807240, S01H003553/000001
Show 3 more code fieldsShow fewer
What the firm is doing
On April 28, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. Circulate this URGENT Medical Device Correction letter to all users of the system so that they are aware of the issue. 2. In case the affected system has been transferred to another organization, please send a copy of this URGENT Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 3. Keep this letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. 4. If you experience the issue described in this letter, please report it to your local Philips representative. Actions planned by Philips IGT Systems to correct the issue Starting in Q1 2027, Philips will replace the deaeration hose in the CU3101 tube cooling unit in all affected systems. Your local Philips representative will contact you to schedule a visit to replace the deaeration hose once available.
DistributionShow detailsHide
Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen;
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2229-2026
- FDA device classification · OWBOfficial FDA classification for this device type
- CFR regulation · 21 CFR 892.1650The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm